1.
J Interv Card Electrophysiol ; 62(1): 103-111, 2021 Oct.
Article in English | MEDLINE | ID: mdl-32965615

ABSTRACT

INTRODUCTION: The clinical impact of slow ventricular tachycardia (VT), occurring in patients carrying implantable cardiac defibrillators (ICD), is still under debate. METHODS AND RESULTS: From the UMBRELLA registry (multicenter, observational, and prospective study on patients with ICD), 659 episodes of slow VT were observed in 97 patients. Untreated slow VT (n = 93) had longer duration (23.7 min, CI95%: 10-39), compared with episodes treated effectively by anti-tachycardia pacing (ATP; n = 527; 0.32 min, IC95%: 0.22-0, 48) or shock (n = 39; 1 min, CI95%: 0.8-1.2). Despite of longer duration, the time to the first contact with the medical services was similar to those episodes treated by ATP (50 days [CI95%: 45-55] vs. 41 days [CI95%: 39-44]). However, both were significantly longer than the time observed in episodes treated with shock (10 days, CI95%: 6-15). This tendency was maintained with successive interrogations of the device (2nd and 3rd). There were no significant differences in mortality during follow-up (48 ± 16 months), neither other adverse outcomes, between patients who presented untreated slow TV and those who did not (log-rank p = 0.28). In a Cox regression analysis, the variable "presenting untreated episodes of slow VT" was not able to predict mortality. However, being in sinus rhythm (vs. atrial fibrillation, OR: 0.31, p = 0.009), narrower QRS (OR: 1.036, p = 0.037) and diabetes (OR 4.673, p = 0.049) appropriately predict survival. CONCLUSIONS: Untreated slow VT does not significantly worsen patient prognosis. Our results support the limitation of therapies to ATP only, thus avoiding therapies that have been associated with increased risk of morbidity and mortality.


Subject(s)
Accelerated Idioventricular Rhythm , Atrial Fibrillation , Defibrillators, Implantable , Tachycardia, Ventricular , Humans , Prospective Studies , Tachycardia, Ventricular/therapy
2.
Europace ; 22(5): 704-715, 2020 05 01.
Article in English | MEDLINE | ID: mdl-31840163

ABSTRACT

AIMS: Atrial electrical remodelling (AER) is a transitional period associated with the progression and long-term maintenance of atrial fibrillation (AF). We aimed to study the progression of AER in individual patients with implantable devices and AF episodes. METHODS AND RESULTS: Observational multicentre study (51 centres) including 4618 patients with implantable cardioverter-defibrillator +/-resynchronization therapy (ICD/CRT-D) and 352 patients (2 centres) with pacemakers (median follow-up: 3.4 years). Atrial activation rate (AAR) was quantified as the frequency of the dominant peak in the signal spectrum of AF episodes with atrial bipolar electrograms. Patients with complete progression of AER, from paroxysmal AF episodes to electrically remodelled persistent AF, were used to depict patient-specific AER slopes. A total of 34 712 AF tracings from 830 patients (87 with pacemakers) were suitable for the study. Complete progression of AER was documented in 216 patients (16 with pacemakers). Patients with persistent AF after completion of AER showed ∼30% faster AAR than patients with paroxysmal AF. The slope of AAR changes during AF progression revealed patient-specific patterns that correlated with the time-to-completion of AER (R2 = 0.85). Pacemaker patients were older than patients with ICD/CRT-Ds (78.3 vs. 67.2 year olds, respectively, P < 0.001) and had a shorter median time-to-completion of AER (24.9 vs. 93.5 days, respectively, P = 0.016). Remote transmissions in patients with ICD/CRT-D devices enabled the estimation of the time-to-completion of AER using the predicted slope of AAR changes from initiation to completion of electrical remodelling (R2 = 0.45). CONCLUSION: The AF progression shows patient-specific patterns of AER, which can be estimated using available remote-monitoring technology.


Subject(s)
Atrial Fibrillation , Atrial Remodeling , Defibrillators, Implantable , Pacemaker, Artificial , Atrial Fibrillation/diagnosis , Atrial Fibrillation/therapy , Child, Preschool , Humans
3.
Eur J Heart Fail ; 20(10): 1472-1481, 2018 10.
Article in English | MEDLINE | ID: mdl-29251799

ABSTRACT

AIMS: We sought to determine whether atrioventricular junction ablation (AVJA) in patients with cardiac resynchronization therapy (CRT) implantable cardioverter-defibrillator (ICD) and with permanent atrial fibrillation (AF) has a positive impact on ICD shocks and hospitalizations compared with rate-slowing drugs. METHODS AND RESULTS: This is a pooled analysis of data from 179 international centres participating in two randomized trials and one prospective observational research. The co-primary endpoints were all-cause ICD shocks and all-cause hospitalizations. Out of 3358 CRT-ICD patients (2720 male, 66.6 years), 2694 (80%) were in sinus rhythm (SR) and 664 (20%) had permanent AF-262 (8%) treated with AVJA (AF + AVJA) and 402 (12%) treated with rate-slowing drugs (AF + Drugs). Median follow-up was 18 months. The mean (95% confidence intervals) annual rate of all-cause ICD shocks per 100 patient years was 8.0 (5.3-11.9) in AF + AVJA, 43.6 (37.7-50.4) in AF + Drugs, and 34.4 (32.5-36.5) in SR patients, resulting in incidence rate ratio (IRR) reductions of 0.18 (0.10-0.32) for AF + AVJA vs. AF + Drugs (P < 0.001) and 0.48 (0.35-0.66) for AF + AVJA vs. SR (P < 0.001). These reductions were driven by significant reductions in both appropriate ICD shocks [IRR 0.23 (0.13-0.40), P < 0.001, vs. AF + Drugs] and inappropriate ICD shocks [IRR 0.09 (0.04-0.21), P < 0.001, vs. AF + Drugs]. Annual rate of all-cause hospitalizations was significantly lower in AF + AVJA vs. AF + Drugs [IRR 0.57 (0.41-0.79), P < 0.001] and SR [IRR 0.85 (073-0.98), P = 0.027]. CONCLUSION: In AF patients treated with CRT, AVJA results in a lower incidence and burden of all-cause, appropriate and inappropriate ICD shocks, as well as to fewer all-cause and heart failure hospitalizations. CLINICAL TRIAL REGISTRATION: NCT00147290, NCT00617175, NCT01007474.


Subject(s)
Atrial Fibrillation/therapy , Atrioventricular Node/surgery , Cardiac Resynchronization Therapy/methods , Catheter Ablation/methods , Defibrillators, Implantable , Hospitalization/trends , Tachycardia, Ventricular/therapy , Aged , Atrioventricular Node/physiopathology , Female , Follow-Up Studies , Heart Rate/physiology , Humans , Male , Prospective Studies , Treatment Outcome
4.
Europace ; 18(11): 1719-1725, 2016 Nov.
Article in English | MEDLINE | ID: mdl-27485577

ABSTRACT

AIMS: A long-detection interval (LDI) (30/40 intervals) has been proved to be superior to a standard-detection interval (SDI) (18/24 intervals) in terms of reducing unnecessary implantable cardioverter defibrillator (ICD) therapies. To better evaluate the different impact of LDI and anti-tachycardia pacing (ATP) on reducing painful shocks, we assessed all treated episodes in the ADVANCE III trial. METHODS AND RESULTS: A total of 452 fast (200 ms < cycle length ≤ 320 ms) arrhythmic episodes were recorded: 284 in 138 patients in the SDI arm and 168 in 82 patients in the LDI arm (106/452 inappropriate detections). A total of 346 fast ventricular tachycardias (FVT) were detected in 169 patients: 208 in 105 patients with SDI and 138 in 64 patients with LDI. Setting LDI determined a significant reduction in appropriate but unnecessary therapies [208 in SDI vs. 138 in LDI; incidence rate ratio (IRR): 0.61 (95% CI 0.45-0.83), P = 0.002]. Anti-tachycardia pacing determined another 52% reduction in unnecessary shocks [208 in SDI with hypothetical shock-only programming vs. 66 in LDI with ATP; IRR: 0.37 (95% CI 0.25-0.53, P < 0.001)]. The efficacy of ATP in terminating FVT was 63% in SDI and 52% in LDI (P = 0.022). No difference in the safety profile (acceleration/degeneration and death/cardiovascular hospitalizations) was observed between the two groups. CONCLUSION: The combination of LDI and ATP during charging is extremely effective and significantly reduces appropriate but unnecessary therapies. The use of LDI alone yielded a 39% reduction in appropriate but unnecessary therapies; ATP on top of LDI determined another 52% reduction in unnecessary shocks. The strategy of associating ATP and LDI could be considered in the majority of ICD recipients.


Subject(s)
Cardiac Pacing, Artificial/methods , Defibrillators, Implantable , Electric Countershock/methods , Tachycardia, Ventricular/therapy , Aged , Female , Humans , Italy , Kaplan-Meier Estimate , Logistic Models , Male , Middle Aged , Secondary Prevention , Single-Blind Method , Tachycardia, Ventricular/diagnosis , Treatment Outcome
5.
Pacing Clin Electrophysiol ; 39(7): 633-41, 2016 Jul.
Article in English | MEDLINE | ID: mdl-27019211

ABSTRACT

BACKGROUND: Different types of ventricular arrhythmias (monomorphic ventricular tachycardia [VT], polymorphic VT, or ventricular fibrillation) can be detected by implantable cardiac defibrillators (ICDs) in fast VT zone. The efficacy of antitachycardia pacing (ATP) depends on the type of the treated arrhythmia. We hypothesized that an automatic algorithm based on morphological affinity of ICD far-field electrograms during tachycardia can predict ATP success and the need of shock. METHODS: The algorithm was evaluated on ventricular arrhythmias recorded in CareLink ICD remote monitoring system (Medtronic Inc., Minneapolis, MN, USA). Patients were selected if first ATP programmed was a burst of eight pulses at 88% coupling interval and if a far-field electrogram was available. The algorithm calculated a stability coefficient (SC) for all their stored ATP-treated fast ventricular arrhythmia (VA) episodes (LC 200-300 ms), analyzing the morphology homogeneity of the last eight recorded far-field electrograms before ventricular arrhythmias detection. RESULTS: Inclusion criteria were fulfilled by 717 patients from 29 centers. Three hundred and twenty fast VA were recorded in 103 patients. A higher SC was observed in episodes terminated with the first-ATP (0.78 [0.72-0.84] vs 0.74 [0.60-0.84]; P = 0.006). These differences were especially marked among the 62 episodes of very fast VA (CL ≤250 ms) (0.77 [0.74-0.85] vs 0.64 [0.51-0.8]; P = 0.006). In the multivariate analysis, a SC > 70% was independently associated with a higher likelihood of first-ATP success (odds ratio [OR] = 2.5; [95% confidence interval (CI) = 1.4-4.5], P = 0.001) and a lower need of shock (OR = 0.37; [95% CI = 0.2-0.7], P = 0.002). CONCLUSION: This automatic algorithm (stability coefficient) shows that ATP therapy response can be predicted in fast ventricular arrhythmias through morphology evaluation.


Subject(s)
Algorithms , Electrophysiologic Techniques, Cardiac/methods , Tachycardia, Ventricular/diagnosis , Tachycardia, Ventricular/prevention & control , Ventricular Fibrillation/diagnosis , Ventricular Fibrillation/prevention & control , Cardiac Pacing, Artificial , Defibrillators, Implantable , Diagnosis, Computer-Assisted/methods , Female , Humans , Male , Middle Aged , Outcome Assessment, Health Care/methods , Prognosis , Reproducibility of Results , Sensitivity and Specificity , Spain , Therapy, Computer-Assisted/methods , Treatment Outcome
6.
Circulation ; 130(4): 308-14, 2014 Jul 22.
Article in English | MEDLINE | ID: mdl-24838360

ABSTRACT

BACKGROUND: Three trials demonstrated recently that a long detection window reduces implantable cardioverter-defibrillator (ICD) therapy in primary prevention patients. Avoid Delivering Therapies for Nonsustained Arrhythmias in ICD Patients III (ADVANCE III) was the only trial that enrolled both primary and secondary prevention patients. METHODS AND RESULTS: Of the 1902 patients enrolled in the ADVANCE III trial, 477 received a defibrillator for secondary prevention; 248 patients were randomly assigned to a long detection setting (30 of 40 intervals) and 229 to the nominal setting (18 of 24 intervals) for ventricular arrhythmias with cycle length ≤ 320 ms. Eight-five percent of patients were men, with a mean age of 65 ± 12 years, a previous history of ventricular fibrillation in 37% of the cases, and a mean ejection fraction of 38 ± 13%. The ICD device mix was 37% single chamber, 47% dual chamber, and 16% triple chamber. Over a median period of 12 months, the long detection period was associated with a 25% reduction in the number of overall therapies (115.6 versus 86.8 per 100 patient-years; incidence rate ratio, 0.75; 95% confidence interval, 0.61-0.93; P=0.008) and a 34% reduction in the number of shocks (rate per 100 patient-years, 51.2 versus 38.1; incidence rate ratio, 0.66; 95% confidence interval, 0.48-0.89; P=0.007). Appropriate therapies (89.7 versus 67.7; incidence rate ratio, 0.77; 95% confidence interval, 0.60-0.97; P=0.029) and appropriate shocks (37.1 versus 28.1; incidence rate ratio, 0.64; 95% confidence interval, 0.45-0.93; P=0.018) were also reduced. CONCLUSIONS: ADVANCE III is the first randomized trial to assess a long detection window setting in ICDs in both primary and secondary prevention populations and demonstrates a reduction of overall therapies and shocks in the subgroup of secondary prevention patients. These data suggest that even the secondary prevention population may benefit from programming that combines a long detection period with antitachycardia pacing during charging. CLINICAL TRIAL REGISTRATION URL: http://www/clinicaltrials.gov. Unique identifier: NCT00617175.


Subject(s)
Defibrillators, Implantable , Electric Countershock/methods , Tachycardia, Ventricular/prevention & control , Ventricular Fibrillation/prevention & control , Aged , Death, Sudden, Cardiac/prevention & control , Electric Countershock/adverse effects , Equipment Failure , Female , Heart Rate , Hospitalization/statistics & numerical data , Humans , Incidence , Male , Middle Aged , Primary Prevention , Secondary Prevention , Single-Blind Method , Stroke Volume , Syncope/epidemiology , Syncope/etiology , Tachycardia, Ventricular/therapy , Time Factors , Ventricular Fibrillation/therapy
SELECTION OF CITATIONS