Dent J (Basel) ; 10(1)2022 Jan 14.
Article in English | MEDLINE | ID: mdl-35049610

ABSTRACT

This study evaluated the influence of distal implants angulation and framework material in the stress concentration of an All-on-4 full-arch prosthesis. A full-arch implant-supported prosthesis 3D model was created with different distal implant angulations and cantilever arms (30° with 10-mm cantilever; 45° with 10-mm cantilever and 45° with 6-mm cantilever) and framework materials (Cobalt-chrome [CoCr alloy], Yttria-stabilized tetragonal zirconia polycrystal [Y-TZP] and polyetheretherketone [PEEK]). Each solid was imported to computer-aided engineering software, and tetrahedral elements formed the mesh. Material properties were assigned to each solid with isotropic and homogeneous behavior. The contacts were considered bonded. A vertical load of 200 N was applied in the distal region of the cantilever arm, and stress was evaluated in Von Misses (σVM) for prosthesis components and the Maximum (σMAX) and Minimum (σMIN) Principal Stresses for the bone. Distal implants angled in 45° with a 10-mm cantilever arm showed the highest stress concentration for all structures with higher stress magnitudes when the PEEK framework was considered. However, distal implants angled in 45° with a 6-mm cantilever arm showed promising mechanical responses with the lowest stress peaks. For the All-on-4 concept, a 45° distal implants angulation is only beneficial if it is possible to reduce the cantilever's length; otherwise, the use of 30° should be considered. Comparing with PEEK, the YTZP and CoCr concentrated stress in the framework structure, reducing the stress in the prosthetic screw.

2.
Dent Mater ; 37(11): 1688-1697, 2021 11.
Article in English | MEDLINE | ID: mdl-34497022

ABSTRACT

OBJECTIVE: The aim of this study was to evaluate the influence of three different dental implant neck geometries, under a combined compressive/shear load using finite element analysis (FEA). The implant neck was positioned in D2 quality bone at the crestal level or 2 mm below. METHODS: One dental implant (4.2 × 9 mm) was digitized by reverse engineering techniques using micro CT and imported into Computer Aided Design (CAD) software. Non-uniform rational B-spline surfaces were reconstructed, generating a 3D volumetric model similar to the digitized implant. Three different models were generated with different implant neck configurations, namely 0°, 10° and 20°. D2 quality bone, composed of cortical and trabecular structure, was modeled using data from CT scans. The implants were included in the bone model using a Boolean operation. Two different fixture insertion depths were simulated for each implant: 2 mm below the crestal bone and exactly at the level of the crestal bone. The obtained models were imported to FEA software in STEP format. Von Mises equivalent strains were analyzed for the peri-implant D2 bone type, considering the magnitude and volume of the affected surrounding cortical and trabecular bone. The highest strain values in both cortical and trabecular tissue at the peri-implant bone interface were extracted and compared. RESULTS: All implant models were able to distribute the load at the bone-implant contact (BIC) with a similar strain pattern between the models. At the cervical region, however, differences were observed: the models with 10° and 20° implant neck configurations (Model B and C), showed a lower strain magnitude when compared to the straight neck (Model A). These values were significantly lower when the implants were situated at crestal bone levels. In the apical area, no differences in strain values were observed. SIGNIFICANCE: The implant neck configuration influenced the strain distribution and magnitude in the cortical bone and cancellous bone tissues. To reduce the strain values and improve the load dissipation in the bone tissue, implants with 10° and 20 neck configuration should be preferred instead of straight implant platforms.


Subject(s)
Dental Implants , Computer-Aided Design , Dental Stress Analysis , Finite Element Analysis , Stress, Mechanical
3.
J Healthc Eng ; 2021: 1342316, 2021.
Article in English | MEDLINE | ID: mdl-33628401

ABSTRACT

[This corrects the article DOI: 10.1155/2020/2707560.].

4.
Materials (Basel) ; 14(1)2021 Jan 02.
Article in English | MEDLINE | ID: mdl-33401673

ABSTRACT

A wide range of materials has been considered to repair cranial defects. In the field of cranioplasty, poly(methyl methacrylate) (PMMA)-based bone cements and modifications through the inclusion of copper doped tricalcium phosphate (Cu-TCP) particles have been already investigated. On the other hand, aliphatic polyesters such as poly(ε-caprolactone) (PCL) and polylactic acid (PLA) have been frequently investigated to make scaffolds for cranial bone regeneration. Accordingly, the aim of the current research was to design and fabricate customized hybrid devices for the repair of large cranial defects integrating the reverse engineering approach with additive manufacturing, The hybrid device consisted of a 3D additive manufactured polyester porous structures infiltrated with PMMA/Cu-TCP (97.5/2.5 w/w) bone cement. Temperature profiles were first evaluated for 3D hybrid devices (PCL/PMMA, PLA/PMMA, PCL/PMMA/Cu-TCP and PLA/PMMA/Cu-TCP). Peak temperatures recorded for hybrid PCL/PMMA and PCL/PMMA/Cu-TCP were significantly lower than those found for the PLA-based ones. Virtual and physical models of customized devices for large cranial defect were developed to assess the feasibility of the proposed technical solutions. A theoretical analysis was preliminarily performed on the entire head model trying to simulate severe impact conditions for people with the customized hybrid device (PCL/PMMA/Cu-TCP) (i.e., a rigid sphere impacting the implant region of the head). Results from finite element analysis (FEA) provided information on the different components of the model.

5.
Polymers (Basel) ; 12(8)2020 Aug 16.
Article in English | MEDLINE | ID: mdl-32824363

ABSTRACT

The aim of the current work was to analyze the influence of the ferrule effect for hybrid composite endodontic post designs consisting of carbon (C) and glass (G) fiber-reinforced polyetherimide (PEI), in upper canine teeth. Starting from theoretical designs of C-G/PEI hybrid composite posts with different Young's moduli (Post A-57.7 GPa, Post B-31.6 GPa, Post C-graduated from 57.7 to 9.0 GPa in the coronal-apical direction) in endodontically treated anterior teeth, the influence of the ferrule effect was determined through finite element analysis (FEA). On the surface of the crown, a load of 50 N was applied at 45° to the longitudinal axis of the tooth. Maximum principal stresses were evaluated along the C-G/PEI post as well as at the interface between the surrounding tooth structure and the post. Maximum stress values were lower than those obtained for the corresponding models without a ferrule. The presence of a ferrule led to a marked decrease of stress and gradients especially for posts A and B. A less marked effect was globally found for Post C, except in a cervical margin section along a specific direction, where a significant decrease of the stress was probably due to local geometric features, compared to the model without a ferrule. The presence of a ferrule did not generally provide a marked benefit in the case of the graduated Post C, in comparison to other C-G/PEI posts. The outcomes suggest how such a hybrid composite post alone should be sufficient to optimize the stress distribution, dissipating stress from the coronal to the apical end.

7.
J Healthc Eng ; 2019: 3212594, 2019.
Article in English | MEDLINE | ID: mdl-31662833

ABSTRACT

Mechanical and architectural features play an important role in designing biomedical devices. The use of materials (i.e., Ti6Al4V) with Young's modulus higher than those of natural tissues generally cause stress shielding effects, bone atrophy, and implant loosening. However, porous devices may be designed to reduce the implant stiffness and, consequently, to improve its stability by promoting tissue ingrowth. If porosity increases, mass transport properties, which are crucial for cell behavior and tissue ingrowth, increase, whereas mechanical properties decrease. As reported in the literature, it is always possible to tailor mass transport and mechanical properties of additively manufactured structures by varying the architectural features, as well as pore shape and size. Even though many studies have already been made on different porous structures with controlled morphology, the aim of current study was to provide only a further analysis on Ti6Al4V lattice structures manufactured by selective laser melting. Experimental and theoretical analyses also demonstrated the possibility to vary the architectural features, pore size, and geometry, without dramatically altering the mechanical performance of the structure.


Subject(s)
Biocompatible Materials/chemistry , Titanium/chemistry , Alloys , Computer Simulation , Elastic Modulus , Lasers , Materials Testing , Porosity
8.
Mater Sci Eng C Mater Biol Appl ; 98: 994-1004, 2019 May.
Article in English | MEDLINE | ID: mdl-30813106

ABSTRACT

The current research reports for the first time the use of blends of poly(ε-caprolactone) (PCL) and poly(ester amide) (PEA) for the fabrication of 3D additive manufactured scaffolds. Tailor made PEA was synthesized to afford fully miscible blends of PCL and PEA using different percentages (5, 10, 15 and 20% w/w). Stability, characteristic temperatures and material's compatibility were studied through thermal analyses (i.e., TGA, DSC). Even though DMTA and static compression tests demonstrated the possibility to improve the storage modulus, Young's modulus and maximum stress by increasing the amount of PEA, a decrease of hardness was found beyond a threshold concentration of PEA as the lowest values were achieved for PCL/PEA (20% w/w) scaffolds (from 0.39 ±â€¯0.03 GPa to 0.21 ±â€¯0.02 GPa in the analysed load range). The scaffolds presented a controlled morphology and a fully interconnected network of internal channels. The water contact angle measurements showed a clear increase of hydrophilicity resulting from the addition of PEA. This result was further corroborated with the improved adhesion and proliferation of human mesenchymal stem cells (hMSCs). The presence of PEA also influenced the cell morphology. Better cell spreading and a much higher and homogenous number of cells were observed for PCL/PEA scaffolds when compared to PCL ones.


Subject(s)
Polyesters/chemistry , Tissue Engineering/methods , Tissue Scaffolds/chemistry , Calorimetry, Differential Scanning , Compressive Strength , Differential Thermal Analysis , Humans , Mesenchymal Stem Cells/cytology , Nanoparticles/chemistry , Polyesters/chemical synthesis , Printing, Three-Dimensional , Stress, Mechanical , Temperature , Thermogravimetry , Water/chemistry
9.
Dent Mater ; 34(8): 1235-1245, 2018 08.
Article in English | MEDLINE | ID: mdl-29891196

ABSTRACT

OBJECTIVE: To investigate the influence of implant design on the change in the natural frequency of bone-implant system during osseointegration by means of a modal 3D finite element analysis. METHODS: Six implants were considered. Solid models were obtained by means of reverse engineering techniques. The mandibular bone geometry was built-up from a CT scan dataset through image segmentation. Each implant was virtually implanted in the mandibular bone. Two different models have been considered, differing in the free length of the mandibular branch ('long branch' and 'short branch') in order to simulate the variability of boundary conditions when performing vibrometric analyses. Modal analyses were carried out for each model, and the first three resonance frequencies were assessed with the respective vibration modes. RESULTS: With reference to the 'long branch' model, the first three modes of vibration are whole bone vibration with minimum displacement of the implant relative to bone, with the exception of the initial condition (1% bone maturation) where the implant is not osseointegrated. By contrast, implant displacements become relevant in the 'short branch' model, unless osseointegration level is beyond 20%. The difference between resonance frequency at whole bone maturation and resonance frequency at 1% bone maturation remained lower than 6.5% for all modes, with the exception of the third mode of vibration in the 'D' implant where this difference reached 9.7%. With reference to the 'short branch', considering the first mode of vibration, 61-68% of the frequency increase was achieved at 10% osseointegration; 72-79% was achieved at 20%; 89-93% was achieved at 50% osseointegration. The pattern of the natural frequency versus the osseointegration level is similar among different modes of vibration. SIGNIFICANCE: Resonance frequencies and their trends towards osseointegration level may differ between implant designs, and in different boundary conditions that are related to implant position inside the mandible; tapered implants are the most sensitive to bone maturation levels, small implants have very little sensitivity. Resonance frequencies are less sensitive to bone maturation level beyond 50%.


Subject(s)
Dental Implants , Dental Prosthesis Design , Dental Prosthesis Retention , Bone-Implant Interface , Computer Simulation , Computer-Aided Design , Dental Materials/chemistry , Dental Stress Analysis/methods , Finite Element Analysis , Osseointegration/physiology , Resonance Frequency Analysis , Tomography, X-Ray Computed , Vibration
10.
Materials (Basel) ; 11(5)2018 May 07.
Article in English | MEDLINE | ID: mdl-29735896

ABSTRACT

A computer-aided design (CAD)-based approach and sol-gel chemistry were used to design a multilayer dental post with a compositional gradient and a Young’s modulus varying from 12.4 to 2.3 GPa in the coronal-apical direction. Specifically, we propose a theoretical multilayer post design, consisting of titanium dioxide (TiO2) and TiO2/poly(ε-caprolactone) (PCL) hybrid materials containing PCL up to 24% by weight obtained using the sol-gel method. The current study aimed to analyze the effect of the designed multilayer dental post in endodontically treated anterior teeth. Stress distribution was investigated along and between the post and the surrounding structures. In comparison to a metal post, the most uniform distributions with lower stress values and no significant stress concentration were found when using the multilayer post.

11.
Eur J Orthod ; 40(6): 592-596, 2018 11 30.
Article in English | MEDLINE | ID: mdl-29726936

ABSTRACT

Objectives: To evaluate the association between maxillary dental arch transverse dimensions, palatal depths, palatal area and volume with buccally displaced canine (BDC) in mixed dentition subjects when compared to non-BDC subjects using laser scanner 3D technology. Materials and methods: Sixty Caucasian subjects, 8-11 years of age (mean, 9.26 ± 1.48 years), were included. In each group (BDC and non-BDC) 30 children were matched. Digital dental casts were obtained using a 3 Shape D700 laser scanner. Intercanine and intermolar widths (cusp and gingival levels), anterior and posterior palatal depth (cusp level), palatal surface area and volume were measured. An independent sample Student's t-test and an ANOVA were undertaken with significance level set as P < 0.05. Results: Intercanine widths at the cusp (1.76 mm; P = 0.020) and the gingival level (1.6 mm; P = 0.006), palatal area (133 mm2; P = 0.021) and volume (790 mm3; P = 0.046) were significantly lower in the BDC compared to the control group. Limitations: A smaller part of the subjects was in late mixed dentition phase. To overcome this limitation a matched control group was used. Some subjects did not have some teeth because of the transition phase which might have had an influence on the dental measurements. However, these subjects were not excluded to avoid introducing a bias. Conclusions: 3D evaluation of the maxillary arch and palate highlighted significant differences between BDC and non-BDC mixed dentition subjects. Maxillary dental arch dimensions and palate morphology may allow early identification and prevention of maxillary canine impaction.


Subject(s)
Cuspid/pathology , Dental Arch/pathology , Palate/pathology , Tooth, Impacted/pathology , Case-Control Studies , Casts, Surgical , Child , Cuspid/diagnostic imaging , Dental Arch/diagnostic imaging , Dentition, Mixed , Female , Humans , Imaging, Three-Dimensional , Male , Maxilla/diagnostic imaging , Maxilla/pathology , Odontometry/methods , Palate/diagnostic imaging , Tooth, Impacted/diagnostic imaging
12.
Dent Mater ; 34(7): 1063-1071, 2018 Jul.
Article in English | MEDLINE | ID: mdl-29703402

ABSTRACT

OBJECTIVES: To assess conceptual designs of dental posts consisting of polyetherimide (PEI) reinforced with carbon (C) and glass (G) glass fibers in endodontically treated anterior teeth. METHODS: 3D tessellated CAD and geometric models of endodontically treated anterior teeth were generated from Micro-CT scan images. Model C-G/PEI composite posts with different Young's moduli were analyzed by Finite Element (FE) methods post A (57.7GPa), post B (31.6GPa), post C (from 57.7 to 9.0GPa in the coronal-apical direction). A load of 50N was applied at 45° to the longitudinal axis of the tooth, acting on the palatal surface of the crown. The maximum principal stress distribution was determined along the post and at the interface between the post and the surrounding structure. RESULTS: Post C, with Young's modulus decreasing from 57.7 to 9.0GPa in the coronal-apical direction, reduced the maximum principal stress distribution in the restored tooth. Post C gave reduced stress and the most uniform stress distribution with no stress concentration, compared to the other C-G/PEI composite posts. SIGNIFICANCE: The FE analysis confirmed the ability of the functionally graded post to dissipate stress from the coronal to the apical end. Hence actual (physical) C-G/PEI posts could permit optimization of stress distributions in endodontically treated anterior teeth.


Subject(s)
Composite Resins/chemistry , Polymers/chemistry , Post and Core Technique , Tooth, Nonvital , Carbon Fiber/chemistry , Computer-Aided Design , Dental Materials/chemistry , Dental Prosthesis Design , Dental Stress Analysis , Elastic Modulus , Finite Element Analysis , Glass/chemistry , Humans , Imaging, Three-Dimensional , In Vitro Techniques , X-Ray Microtomography
13.
Materials (Basel) ; 11(4)2018 Mar 27.
Article in English | MEDLINE | ID: mdl-29584683

ABSTRACT

Over the last three decades, it has been frequently reported that the properties of dental restorative composites cured with argon laser are similar or superior to those achieved with conventional halogen and light emitting diode (LED) curing units. Whereas laser curing is not dependent on the distance between the curing unit and the material, such distance represents a drawback for conventional curing units. However, a widespread clinical application of this kind of laser remains difficult due to cost, heavy weight, and bulky size. Recently, with regard to the radiation in the blue region of the spectrum, powerful solid-state lasers have been commercialized. In the current research, CAD (computer-aided design)/CAM (computer-aided manufacturing) assisted solid-state lasers were employed for curing of different dental restorative composites consisting of micro- and nanoparticle-reinforced materials based on acrylic resins. Commercial LED curing units were used as a control. Temperature rise during the photopolymerisation process and bending properties were measured. By providing similar light energy dose, no significant difference in temperature rise was observed when the two light sources provided similar intensity. In addition, after 7 days since curing, bending properties of composites cured with laser and LED were similar. The results suggested that this kind of laser would be suitable for curing dental composites, and the curing process does not suffer from the tip-to-tooth distance.

14.
Materials (Basel) ; 11(2)2018 Feb 21.
Article in English | MEDLINE | ID: mdl-29466299

ABSTRACT

Experimental/theoretical analyses have already been performed on poly(ε-caprolactone) (PCL) loaded with organic-inorganic fillers (PCL/TiO2 and PCL/ZrO2) to find a correlation between the results from the small punch test and Young's modulus of the materials. PCL loaded with Ti2 (PCL = 12, TiO2 = 88 wt %) and Zr2 (PCL = 12, ZrO2 = 88 wt %) hybrid fillers showed better performances than those obtained for the other particle composition. In this context, the aim of current research is to provide further insight into the mechanical properties of PCL loaded with sol-gel-synthesized organic-inorganic hybrid fillers for bone tissue engineering. For this reason, theoretical analyses were performed by the finite element method. The results from the small punch test and Young's modulus of the materials were newly correlated. The obtained values of Young's modulus (193 MPa for PCL, 378 MPa for PCL/Ti2 and 415 MPa for PCL/Zr2) were higher than those obtained from a previous theoretical modelling (144 MPa for PCL, 282 MPa for PCL/Ti2 and 310 MPa for PCL/Zr2). This correlation will be an important step for the evaluation of Young's modulus, starting from the small punch test data.

15.
Dent Mater ; 34(3): 460-469, 2018 03.
Article in English | MEDLINE | ID: mdl-29395470

ABSTRACT

OBJECTIVE: To assess the influence of implant thread shape and inclination on the mechanical behaviour of bone-implant systems. The study assesses which factors influence the initial and full osseointegration stages. METHODS: Point clouds of the original implant were created using a non-contact reverse engineering technique. A 3D tessellated surface was created using Geomagic Studio® software. From cross-section curves, generated by intersecting the tessellated model and cutting-planes, a 3D parametric CAD model was created using SolidWorks® 2017. By the permutation of three thread shapes (rectangular, 30° trapezoidal, 45° trapezoidal) and three thread inclinations (0°, 3° or 6°), nine geometric configurations were obtained. Two different osseointegration stages were analysed: the initial osseointegration and a full osseointegration. In total, 18 different FE models were analysed and two load conditions were applied to each model. The mechanical behaviour of the models was analysed by Finite Element (FE) Analysis using ANSYS® v. 17.0. Static linear analyses were also carried out. RESULTS: ANOVA was used to assess the influence of each factor. Models with a rectangular thread and 6° inclination provided the best results and reduced displacement in the initial osseointegration stages up to 4.58%. This configuration also reduced equivalent VM stress peaks up to 54%. The same effect was confirmed for the full osseointegration stage, where 6° inclination reduced stress peaks by up to 62%. SIGNIFICANCE: The FE analysis confirmed the beneficial effect of thread inclination, reducing the displacement in immediate post-operative conditions and equivalent VM stress peaks. Thread shape does not significantly influence the mechanical behaviour of bone-implant systems but contributes to reducing stress peaks in the trabecular bone in both the initial and full osseointegration stages.


Subject(s)
Bone-Implant Interface/physiology , Dental Implants , Dental Prosthesis Design , Osseointegration/physiology , Biomechanical Phenomena , Computer Simulation , Finite Element Analysis , Software
16.
Am J Dent ; 30(4): 227-232, 2017 Aug.
Article in English | MEDLINE | ID: mdl-29178706

ABSTRACT

PURPOSE: To study the influence of the resin bonding layer thickness and the bulk filling material stiffness in adhesive class II mesio-occlusal-distal (MOD) restorations using numerical finite element analysis (FEA). METHODS: Four 3D-FE models of teeth restored with different filling material stiffness and resin bonding layer thickness were built-up and analyzed. The 3D model of a sound lower molar was also analyzed and compared with restored ones. The tooth tissues (enamel, dentin), dental restoration and bolus on the occlusal surface, was divided into 3D solid CTETRA elements with four grid points. The adhesive bonding around the dental restoration was modeled with shell elements. Polymerization shrinkage was simulated with a thermal expansion approach. Mechanical behavior of restored models in terms of stress and displacement distributions, under the combination effects of polymerization shrinkage and occlusal load (600 N), was analyzed. All the materials were assumed to behave as elastic materials throughout the entire deformation. RESULTS: Numerical results show that the mechanical response of the restored models was very different compared to the sound tooth ones, where the stress was uniformly distributed from enamel to dentin with no critical stress concentration. In the restored models, the highest stress values were detected in the enamel, near the enamel-dentin interface and in the bulk restorative material. Tooth preparations A and B showed lower gradient stresses than corresponding C and D. The value of the vertical displacement components in models A and B were higher than corresponding C and D. The maximum displacement values were mainly located around the groove and were higher by an order of magnitude than the sound models. The results showed better mechanical response with models A and B compared to C and D. It is also evident that resin bonding thickness slightly affected the stress level of the restored teeth. CLINICAL SIGNIFICANCE: Class II MOD direct bulk resin composite restorations showed a high susceptibility to damage at the marginal and internal tissue interfaces depending on their own stiffness. The use of resin-based bulk filling materials is not recommended for large class II MOD adhesive restorations due to mechanical behavior failure risk.


Subject(s)
Composite Resins , Dental Restoration, Permanent , Dental Cavity Preparation , Dental Materials , Dental Stress Analysis , Finite Element Analysis , Materials Testing , Stress, Mechanical
17.
Dent Mater ; 33(12): 1466-1472, 2017 12.
Article in English | MEDLINE | ID: mdl-29126633

ABSTRACT

OBJECTIVE: To assess the effect of a ferrule design with specific post material-shape combinations on the mechanical behavior of post-restored canine teeth. METHODS: Micro-CT scan images of an intact canine were used to create a 3-D tessellated CAD model, from which the shapes of dentin, pulp and enamel were obtained and geometric models of post-endodontically restored teeth were created. Two types of 15mm post were evaluated: a quartz fiber post with conical-tapered shape, and a carbon (C) fiber post with conical-cylindrical shape. The abutment was created around the coronal portion of the posts and 0.1mm cement was added between prepared crown and abutment. Cement was also added between the post and root canal and a 0.25mm periodontal ligament was modeled around the root. Four models were analysed by Finite Element (FE) Analysis: with/without a ferrule for both types of post material and shape. A load of 50N was applied at 45° to the longitudinal axis of the tooth, acting on the palatal surface of the crown. The maximum normal stress criterion was adopted as a measure of potential damage. RESULTS: Models without a ferrule showed greater stresses (16.3MPa) than those for models with a ferrule (9.2MPa). With a ferrule, stress was uniformly distributed along the abutment and the root, with no critical stress concentration. In all models, the highest stresses were in the palatal wall of the root. Models with the C-fiber post had higher stress than models with the quartz fiber posts. The most uniform stress distribution was with the combination of ferrule and quartz fiber post. SIGNIFICANCE: The FE analysis confirmed a beneficial ferrule effect with the combination of ferrule and quartz fiber post, with tapered shape, affording no critical stress concentrations within the restored system.


Subject(s)
Crowns , Cuspid/physiology , Root Canal Therapy , Tooth, Nonvital/physiopathology , Computer-Aided Design , Cuspid/diagnostic imaging , Dental Abutments , Dental Materials/chemistry , Dental Prosthesis Design , Finite Element Analysis , Humans , Imaging, Three-Dimensional , Materials Testing , Post and Core Technique , Tooth, Nonvital/diagnostic imaging , X-Ray Microtomography
18.
Dent Mater ; 33(12): 1456-1465, 2017 12.
Article in English | MEDLINE | ID: mdl-29128128

ABSTRACT

OBJECTIVES: To investigate the influence of specific resin-composite, glass ceramic and glass ionomer cement (GIC) material combinations in a "multi-layer" technique to replace enamel and dentin in class II mesio-occlusal-distal (MOD) dental restorations using 3D-Finite Element Analysis (FEA). METHODS: Four 3D-FE models (A-D) of teeth, adhesively restored with different filling materials, were created and analyzed in comparison with a 3D model (E) of a sound lower molar. Models A, B & C had "multilayer" constructions, consisting of three layers: adhesive, dentin replacement and enamel replacement. Model A: had a low modulus (8GPa) composite replacing dentin and a higher modulus (12GPa) composite replacing enamel. Model B: had a GI cement replacing dentin and a higher modulus (12GPa) composite replacing enamel. Model C: had a low modulus (8GPa) composite replacing dentin and a very high modulus (70GPa) inlay replacing enamel. Model D: had a lithium disilicate inlay replacing both dentin and enamel with a luting cement base-layer. Polymerization shrinkage effects were simulated and a load of 600N was applied. All the materials were assumed to behave elastically throughout the entire deformation. RESULTS: Model A showed the highest stress distribution along all the adhesive interfaces of the shrinking resin-based materials with a critical condition and failure risk marginally and internally. Model D, by contrast, showed a more favorable performance than either of the multilayer groups (A-C). Stress and displacement plots showed an elastic response similar to that obtained for the sound tooth model. Model B and Model C performed according to their bilayer material properties. The use of a non-shrink dentin component simulating a GIC clearly affected the shrinkage stress at the basis of the Model B; while the bulk resin composite having a 12GPa Young's modulus and linear polymerization shrinkage of 1% strongly influenced the biomechanical response in the bucco-lingual direction. SIGNIFICANCE: Direct resin-based composite materials applied in multilayer techniques to large class II cavities, with or without shrinking dentin layers, produced adverse FEA stress distributions and displacements. An indirect lithium disilicate inlay used to replace lost dentin and enamel in posterior restored teeth generated lower stress levels, within the limits of the elastic FEA model.


Subject(s)
Ceramics/chemistry , Composite Resins/chemistry , Dental Materials/chemistry , Dental Restoration, Permanent/methods , Glass Ionomer Cements/chemistry , Computer Simulation , Computer-Aided Design , Dental Stress Analysis , Finite Element Analysis , Humans , Materials Testing , Molar , X-Ray Microtomography
19.
Materials (Basel) ; 10(2)2017 Feb 08.
Article in English | MEDLINE | ID: mdl-28772505

ABSTRACT

A statistical approach for the characterization of Additive Manufacturing (AM) processes is presented in this paper. Design of Experiments (DOE) and ANalysis of VAriance (ANOVA), both based on Nested Effects Modeling (NEM) technique, are adopted to assess the effect of different laser exposure strategies on physical and mechanical properties of AlSi10Mg parts produced by Direct Metal Laser Sintering (DMLS). Due to the wide industrial interest in AM technologies in many different fields, it is extremely important to ensure high parts performances and productivity. For this aim, the present paper focuses on the evaluation of tensile properties of specimens built with different laser exposure strategies. Two optimal laser parameters settings, in terms of both process quality (part performances) and productivity (part build rate), are identified.

20.
Dent Mater ; 33(6): 690-701, 2017 06.
Article in English | MEDLINE | ID: mdl-28413061

ABSTRACT

OBJECTIVES: To study the influence of resin based and lithium disilicate materials on the stress and strain distributions in adhesive class II mesio-occlusal-distal (MOD) restorations using numerical finite element analysis (FEA). To investigate the materials combinations in the restored teeth during mastication and their ability to relieve stresses. METHODS: One 3D model of a sound lower molar and three 3D class II MOD cavity models with 95° cavity-margin-angle shapes were modelled. Different material combinations were simulated: model A, with a 10µm thick resin bonding layer and a resin composite bulk filling material; model B, with a 70µm resin cement with an indirect CAD-CAM resin composite inlay; model C, with a 70µm thick resin cement with an indirect lithium disilicate machinable inlay. To simulate polymerization shrinkage effects in the adhesive layers and bulk fill composite, the thermal expansion approach was used. Shell elements were employed for representing the adhesive layers. 3D solid CTETRA elements with four grid points were employed for modelling the food bolus and tooth. Slide-type contact elements were used between the tooth surface and food. A vertical occlusal load of 600 N was applied, and nodal displacements on the bottom cutting surfaces were constrained in all directions. All the materials were assumed to be isotropic and elastic and a static linear analysis was performed. RESULTS: Displacements were different in models A, B and C. Polymerization shrinkage hardly affected model A and mastication only partially affected mechanical behavior. Shrinkage stress peaks were mainly located marginally along the enamel-restoration interface at occlusal and mesio-distal sites. However, at the internal dentinal walls, stress distributions were critical with the highest maximum stresses concentrated in the proximal boxes. In models B and C, shrinkage stress was only produced by the 70µm thick resin layer, but the magnitudes depended on the Young's modulus (E) of the inlay materials. Model B mastication behavior (with E=20GPa) was similar to the sound tooth stress relief pattern. Model B internally showed differences from the sound tooth model but reduced maximum stresses than model A and partially than model C. Model C (with E=70GPa) behaved similarly to model B with well redistributed stresses at the occlusal margins and the lateral sides with higher stress concentrations in the proximal boxes. Models B and C showed a more favorable performance than model A with elastic biomechanics similar to the sound tooth model. SIGNIFICANCE: Bulk filling resin composite with 1% linear polymerization shrinkage negatively affected the mechanical behavior of class II MOD restored teeth. Class II MOD direct resin composite showed greater potential for damage because of higher internal and marginal stress evolution during resin polymerization shrinkage. With a large class II MOD cavity an indirect composite or a lithium disilicate inlay restoration may provide a mechanical response close to that of a sound tooth.


Subject(s)
Dental Porcelain , Dental Restoration, Permanent , Composite Resins , Finite Element Analysis , Materials Testing , Stress, Mechanical
SELECTION OF CITATIONS