Talanta ; 58(6): 1109-14, 2002 Dec 06.
Article in English | MEDLINE | ID: mdl-18968846

ABSTRACT

A potentiometric SI titration system for the determination of total acidity in soft drinks is proposed. The concept is based on the aspiration of the acid soft drink sample between two base zones into a holding coil with the volume of the first base zone twice to that of the second one and channelled by flow reversal through a reaction coil to a potentiometric sensor. A solution of 0.1 mol l(-1) sodium chloride is used as ionic strength adjustment buffer in the carrier stream. The system has been applied to the analysis of some South African soft drinks having a total acidity level of about 0.2-0.3% (w/v). The method has a sample frequency of 45 samples per h with a linear range of 0.1 and 0.6% (w/v). It is easy to use, fully computerised, and gives the results that are comparable to both automated batch titration and manual titration.