1.
Clin Phys Physiol Meas ; 12 Suppl B: 21-9, 1991.
Article in English | MEDLINE | ID: mdl-1807875

ABSTRACT

A 37-channel DC SQUID magnetometer system has been built for biomagnetic studies. The SQUID loop of each magnetometer serves as the active sensing element, thereby eliminating the need for flux coupling circuits. The magnetometers are located approximately 3 cm above the outer dewar bottom. The SQUIDs are directly coupled to highly simplified read-out electronics using only five wires per channel; no helium temperature impedance matching circuits are required. Each channel can be independently inserted into or removed from the dewar. Using a novel electronic noise reduction technique the system white and 1 Hz flux density noise values are typically 5 and 10 fT Hz-1/2, respectively, including the noise contribution of the in-house fabricated dewar and the magnetically shielded room. The two parts of the data-processing system allow independent handling of the acquisition and analysing task. Two example measurements demonstrate the advantage of the electronic noise reduction method.


Subject(s)
Magnetics , Equipment Design , Humans