Health Phys ; 114(5): 543-548, 2018 05.
Article in English | MEDLINE | ID: mdl-29578903

ABSTRACT

The U.S. Navy uses the Harshaw 8840/8841 dosimetric (DT-702/PD) system, which employs LiF:Mg,Cu,P thermoluminescent dosimeters (TLDs), developed and produced by Thermo Fisher Scientific (TFS). The dosimeter consists of four LiF:Mg,Cu,P elements, mounted in Teflon® on an aluminum card and placed in a plastic holder. The holder contains a unique filter for each chip made of copper, acrylonitrile butadiene styrene (ABS), Mylar®, and tin. For accredited dosimetry labs, the ISO/IEC 17025:2005(E) requires an acceptance procedure for all new equipment. The Naval Dosimetry Center (NDC) has developed and tested a new non-destructive procedure, which enables the verification and the evaluation of embedded filters in the holders. Testing is based on attenuation measurements of low-energy radiation transmitted through each filter in a representative sample group of holders to verify that the correct filter type and thickness are present. The measured response ratios are then compared with the expected response ratios. In addition, each element's measured response is compared to the mean response of the group. The test was designed and tested to identify significant nonconformities, such as missing copper or tin filters, double copper or double tin filters, or other nonconformities that may impact TLD response ratios. During the implementation of the developed procedure, testing revealed a holder with a double copper filter. To complete the evaluation, the impact of the nonconformities on proficiency testing was examined. The evaluation revealed failures in proficiency testing categories III and IV when these dosimeters were irradiated to high-energy betas.


Subject(s)
Occupational Exposure/analysis , Radiation Dosimeters/standards , Radiation Protection/instrumentation , Technology Assessment, Biomedical/methods , Thermoluminescent Dosimetry/methods , Calibration , Equipment Design , Humans , Military Personnel , Radiation Dosage , Technology Assessment, Biomedical/standards , Thermoluminescent Dosimetry/instrumentation , Thermoluminescent Dosimetry/standards , United States
SELECTION OF CITATIONS
SEARCH DETAIL
...