Lett Appl Microbiol ; 57(4): 282-7, 2013 Oct.
Article in English | MEDLINE | ID: mdl-23710651

ABSTRACT

UNLABELLED: A repeatable and sensitive method to evaluate the effect of three antiseptics and two disinfection techniques on viable micro-organisms on luer-activated catheter needleless connectors (NCs) was developed. NCs were inoculated with Staphylococcus epidermidis or Klebsiella pneumoniae and disinfected with 3·15% chlorhexidine gluconate + 70% isopropanol (CGI), 70% isopropanol (IPA) or 10% PVP povidone-iodine (PI) antiseptic pads using: (i) scrubbing the NC septum and threaded external surfaces or (ii) wiping only the surface of the septum. Treatments were also evaluated against NCs pretreated with human serum and exposed for 18 h to Staph. epidermidis prior to testing. Viable cells were quantified by plate count. The method for inoculation and recovery of luminal micro-organisms was repeatable (SD, 0·31; n = 28). IPA disinfection provided an approximate 3 log10  CFU reduction; CGI and PI provided 3-4 log10 reductions. PI and CGI were more effective than IPA (P < 0·05), but differences between CGI and PI were not significant for either disinfection method. IPA, but not CGI and PI was also less effective (P < 0·05) against NCs inoculated with Kl. pneumoniae than Staph. epidermidis. Pretreatment with serum and prolonged Staph. epidermidis inoculation removed the advantage seen with CGI and PI; log10 reductions were 1·80, 1·73 and 2·50 for CGI, PI and IPA, respectively. PI or CGI may be more effective than IPA for NC disinfection but effectiveness may be reduced on NCs contaminated with blood or serum. SIGNIFICANCE AND IMPACT OF THE STUDY: sensitive and repeatable protocol was developed to evaluate antiseptics for disinfecting catheter needleless connectors (NCs). Povidone-iodine (PI) and chlorhexidine gluconate plus isopropanol (CGI) were more effective than isopropanol (IPA) for reducing Staphylococcus epidermidis contamination of NCs. The effectiveness of PI and CGI was reduced on NCs pre-exposed to human serum and prolonged bacterial inoculation. IPA was also less effective against NCs contaminated with Klebsiella pneumoniae.


Subject(s)
Catheter-Related Infections/prevention & control , Central Venous Catheters/microbiology , Disinfectants/pharmacology , Disinfection/methods , Equipment Contamination/prevention & control , Catheter-Related Infections/microbiology , Chlorhexidine/analogs & derivatives , Chlorhexidine/pharmacology , Disinfection/instrumentation , Humans , Klebsiella pneumoniae/drug effects , Povidone-Iodine/pharmacology , Staphylococcus epidermidis/drug effects
SELECTION OF CITATIONS
SEARCH DETAIL
...