Skeletal Radiol ; 30(7): 398-401, 2001 Jul.
Article in English | MEDLINE | ID: mdl-11499781

ABSTRACT

OBJECTIVE: To quantify image artifact reduction using a new technique (MARS--metal artifact reduction sequence) in vitro. DESIGN: Coronal T1-weighted MR images were obtained through two metal phantoms (titanium/chromium-cobalt and stainless steel femoral prostheses) immersed in water. Comparison of artifact volume was made with images obtained using conventional and modified (MARS) T1-weighted sequences. Signal intensity values outside a range of +/-40% the average signal intensity for water were considered artifact and segmented into low or high signal artifact categories. Considering the arbitrary selection of this threshold value, volumetric calculations of artifact were also evaluated at +/-50%, 60%, 70%, and 80% the mean signal for water. RESULTS: Conventional T1-weighted images produced 87% more low signal artifact and 212% more high signal artifact compared with the MARS modified T1-weighted images of the stainless steel prosthesis. Conventional T1-weighted images of the titanium prosthesis produced 84% more low signal artifact and 211% more high signal artifact than the MARS modified sequence. The level of artifact reduction was essentially uniform for the various threshold levels tested and was greatest at +/-20% the global signal intensity average for water. CONCLUSION: The MARS technique reduces the volume of image signal artifact produced by stainless steel and titanium/chromium-cobalt femoral prostheses on T1-weighted spin-echo images in a tissue phantom model.


Subject(s)
Artifacts , Magnetic Resonance Imaging/methods , Metals , Chromium , Cobalt , Hip Prosthesis , Humans , Magnetic Resonance Imaging/instrumentation , Phantoms, Imaging , Prostheses and Implants , Stainless Steel , Titanium
2.
Skeletal Radiol ; 25(5): 425-30, 1996 Jul.
Article in English | MEDLINE | ID: mdl-8837273

ABSTRACT

We present the MR appearances of three patients with biopsy-proven primary lymphoma of skeletal muscle. In each case lymphoma resulted in bulky expansion of the involved muscle, homogeneously isointense to skeletal muscle on T1-weighted images, homogeneously hyperintense to skeletal muscle on T2-weighted images and diffusely enhancing following intravenous administration of gadopentate dimeglumine.


Subject(s)
Lymphoma, B-Cell/pathology , Magnetic Resonance Imaging , Muscle Neoplasms/pathology , Muscle, Skeletal/pathology , Adult , Aged , Female , Humans , Lymphoma, Follicular/pathology , Lymphoma, Large B-Cell, Diffuse/pathology , Male , Retrospective Studies