J Oral Sci ; 54(3): 227-32, 2012 Sep.
Article in English | MEDLINE | ID: mdl-23047033

ABSTRACT

Each of three cone beam computed tomography (CBCT) machines offer a small field of view (FOV); Promax, Accuitomo 3D, and Kodak 9000. The objectives were to determine, when using a small FOV, the difference in contrast-to-noise ratios (CNR) of images obtained from different machines and different scan settings within each machine. A phantom made of a dry mandible with an epoxy resin based substitute (ERBS) block was used. The ERBS block was fixed to the lingual cortex in the posterior left mandible. Water was used to simulate soft tissue. Twenty scans were acquired. CNR was calculated as the difference in density of the ERBS block and a control area divided by the standard deviation of the difference. The CNR obtained from images acquired with the Promax machine (voxel size of 0.2 mm) was higher than the other machines (voxel sizes 0.076, 0.08, or 0.125 mm) when using a small FOV. The Accuitomo 360° scan had higher CNR than the Accuitomo 180° scan due to the doubling of the number of basis images acquired. Smaller voxel size does not necessarily lead to enhanced CNR. Increasing the number of basis images did increase the CNR.


Subject(s)
Cone-Beam Computed Tomography/instrumentation , Tomography Scanners, X-Ray Computed , Cone-Beam Computed Tomography/methods , Humans , Mandible/diagnostic imaging , Phantoms, Imaging , Signal-To-Noise Ratio