ABSTRACT

This work demonstrates the use of high-resolution 3D printing to fine-tune the low energy dependence of an eye lens dosimeter holder associated to a BeO OSL detector element (ezClip). Five geometries of the denominated iBe dosimeter were developed, three with a variation in the thickness of the wall in front of the sensitive element that tailor the response at low radiation energies; and three with variations of width and curvature in order to vary the angular response of the dosimeter badges. Additive manufacturing was accomplished using stereolithography which gave a high degree of accuracy and precision. The optimised dosimeter badges showed a low energy and angular dependence, within -20% to +20% in the energy range of 24 keV to 662 keV and from 0 to 60° incidence; and within -10% to +10% in the energy range of 24 keV to 164 keV and from 0 to 60° incidence. In contrast to other dosimeters with higher effective atomic numbers, the use of BeO as the sensitive element resulted in a flat energy and angular dependence response at low energies. A significant reduction in the measurement uncertainty in the diagnostic radiology energy range was achieved.


Subject(s)
Lens, Crystalline , Radiation Dosimeters , Photons , Printing, Three-Dimensional , Radiometry