Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters










Database
Language
Publication year range
1.
AAPS J ; 17(6): 1511-6, 2015 Nov.
Article in English | MEDLINE | ID: mdl-26139446

ABSTRACT

Immunogenicity testing for PEGylated biotherapeutics should include methods to detect both anti-protein and anti-PEG antibodies (anti-PEG). Although some methods have been published for the detection of anti-PEG antibodies, the information is incomplete and, in some cases, reagents used (such as Tween-20) are known to interfere with detection. This rapid communication describes the use of BioScale's Acoustic Membrane MicroParticle (AMMP®) technology using the ViBE® Workstation to measure anti-PEG antibodies in human serum samples. Briefly, a sample spiked with monoclonal human IgG anti-PEG antibody is diluted in buffer and incubated with paramagnetic beads coated with linear chain mPEG to capture anti-PEG antibodies. The complex is then captured on an acoustic membrane coated with Protein A. The change in mass on the membrane caused by the binding of the complex to the membrane results in a signal proportional to the mass of anti-PEG antibodies. The data indicate that an assay with a sensitivity of less than 1000 ng/mL for IgG is achievable. This level of sensitivity is better than current published reports on IgG anti-PEG antibody detection.


Subject(s)
Antibodies, Anti-Idiotypic/blood , Chemistry, Pharmaceutical/methods , Immunoglobulin G/blood , Polyethylene Glycols/analysis , Biotin/blood , Enzyme-Linked Immunosorbent Assay/methods , Humans
SELECTION OF CITATIONS
SEARCH DETAIL
...