Anaesth Intensive Care ; 24(4): 472-6, 1996 Aug.
Article in English | MEDLINE | ID: mdl-8862645

ABSTRACT

This study compared two bedside methods recommended for the detection of low concentrations of heparin and the activated partial thromboplastin time (APTT), with reference to a laboratory measure of heparin concentration. Patients undergoing cardiopulmonary bypass had blood drawn at four stages when low levels of heparin could be expected. At each stage four tests were performed: whole blood clotting time using a Hemochron analyser with a Saline-Rinsed test cartridge, whole blood clotting time using a Hemotec analyser with a High Range Heparinase test cartridge, APTT, and heparin concentration by polybrene neutralization. Thirty patients were studied. The sensitivity of the Saline-Rinsed Hemochron, Hemotec High Range Heparinase, and APTT in detecting concentrations of heparin less than 1 U/ml was 38%, 40% and 97%, respectively, while specificities were 87%, 90%, and 30%, respectively. Neither the Saline Rinsed Hemochron, nor the Hemotec Heparinase cartridge reliably detected concentrations of heparin less than 1 U/ml.


Subject(s)
Anticoagulants/blood , Equipment and Supplies , Heparin/blood , Partial Thromboplastin Time , Anticoagulants/administration & dosage , Cardiopulmonary Bypass , Evaluation Studies as Topic , Heparin/administration & dosage , Heparin/deficiency , Humans , Infusions, Intravenous , Middle Aged , Preoperative Care , Sensitivity and Specificity