Med Tekh ; (2): 6-8, 1992.
Article in Russian | MEDLINE | ID: mdl-1513245

ABSTRACT

The authors describe a device for measuring the dose field in the beam section of ionizing radiation. The device consists of an immovable bench mark detector and the two detectors: a movable one and a photodiode placed on a tele- and separately movable carriage. The device affords defining the correspondence of radiation and light fields.


Subject(s)
Radiation Dosage , Radiometry/instrumentation , Radiation, Ionizing
4.
Med Tekh ; (6): 10-2, 1990.
Article in Russian | MEDLINE | ID: mdl-2290368

ABSTRACT

A study was made of the dosimetric characteristics of semiconductor detectors (on the basis of silicon with the p-i-n structure) and combined detectors (scintillation crystals based on CsJ/Tl combined with silicon detectors). The spectral characteristics of semiconductor and combined detectors are described. The results of levelling of hardness dependence by filtration and combination of detectors are presented; their isotropism is reviewed.


Subject(s)
Radiotherapy Dosage , Humans , Radiation , Semiconductors , Sunlight
5.
Med Tekh ; (1): 41-2, 1990.
Article in Russian | MEDLINE | ID: mdl-2338912

ABSTRACT

Three varieties of complete control over dosimetric devices are reviewed. In these devices, use is made of combined detectors consisting of scintillators and semi-conductors. The control is effected by means of scintillator illumination from a stabilized light source via an edge-lighted panel.


Subject(s)
Radiometry/instrumentation
6.
Med Radiol (Mosk) ; 34(4): 11-6, 1989 Apr.
Article in Russian | MEDLINE | ID: mdl-2496282

ABSTRACT

The authors describe optimum plans of irradiation of central lung cancer of patients with different stature in a restricted number of radiation directions and without such restrictions. The recommendations obtained are compared with routine methods of irradiation. The description of a single dose field by transformed curves of axial distribution, taking into account the heterogeneity of a medium, are used in computation.


Subject(s)
Lung Neoplasms/radiotherapy , Radiotherapy, High-Energy/methods , Humans , Particle Accelerators , Radiotherapy Planning, Computer-Assisted
8.
Med Radiol (Mosk) ; 33(8): 51-3, 1988 Aug.
Article in Russian | MEDLINE | ID: mdl-3412140

ABSTRACT

Analysis of dose fields in a heterogeneous tissue equivalent medium has shown that dose distributions have radial symmetry and can be described by a curve of axial distribution with renormalization of maximum ionization depth. A method of the calculation of a dose field in a heterogeneous medium using the principle of radial symmetry is presented.


Subject(s)
Radiation Dosage , Algorithms , Bone and Bones/radiation effects , Humans , Lung/radiation effects , Models, Biological
9.
Med Tekh ; (4): 49-51, 1987.
Article in Russian | MEDLINE | ID: mdl-3670017

ABSTRACT

The paper describes a digital apparatus capable of measuring the duration of X-radiation within the range of 0.001-10.0 s with an error of 2%. The apparatus is used for exposure control in X-ray units having four- and six-pulse circuits.


Subject(s)
Radiography/instrumentation , Humans
10.
Med Radiol (Mosk) ; 31(7): 27-32, 1986 Jul.
Article in Russian | MEDLINE | ID: mdl-3736384

ABSTRACT

The method of treatment of uterine cancer by the brake irradiation of 25 MeV betatron using original devices which promote forming therapeutic figured bunches is presented. The binding of the protective blocks with a special adjusting frame within the aperture of the diaphragm provided for low relative entering dose which is the advantage of high energy irradiation bunch. The use of the forming devices makes it possible to practice individual treatment and decrease the levels of irradiation doses for intact organs and tissues.


Subject(s)
Uterine Neoplasms/radiotherapy , Electrons , Female , Humans , Radiation Protection , Radiotherapy Dosage , Uterine Cervical Neoplasms/radiotherapy
12.
Med Tekh ; (6): 26-9, 1984.
Article in Russian | MEDLINE | ID: mdl-6530941

ABSTRACT

Dosimetry systems for evaluating radiation parameters over the X-ray diagnostic range and for documenting patient exposure doses are analyzed. In the general case the hardware for recording the complete set of parameters is proposed, and simplified versions for particular problems are also given.


Subject(s)
Radiography/standards , Humans , Radiation Dosage , Radiometry/instrumentation
13.
Med Radiol (Mosk) ; 29(10): 23-6, 1984 Oct.
Article in Russian | MEDLINE | ID: mdl-6503637

ABSTRACT

Combined therapy with accelerator high-energy radiations was applied to 116 patients with breast cancer, Stages II-III. At the beginning the patients received preoperative irradiation of the affected breast and ipsilateral axillary zone with braking irradiation at 25 MeV. The summary focal dose of 20 Gy was delivered with 5 daily fractions, 4 Gy each. This was followed by radical mastectomy. In the postoperative period irradiation of the supraclavicular and parasternal zones on the ipsilateral side with electron radiation at 10-22 McV was used when indicated. The summary focal dose of 44-50 Gy was provided in 22-25 fractions. Local recurrences developed in 2.6% of the cases. No regional recurrences and metastases to the supraclavicular lymph nodes were noted. No marked reactions and complications were revealed.


Subject(s)
Breast Neoplasms/radiotherapy , Axilla , Breast Neoplasms/mortality , Breast Neoplasms/surgery , Combined Modality Therapy , Electrons , Female , Follow-Up Studies , Humans , Lymphatic Metastasis , Mastectomy , Neoplasm Recurrence, Local , Radiation Injuries , Radiotherapy Dosage
15.
Med Tekh ; (4): 54-6, 1984.
Article in Russian | MEDLINE | ID: mdl-6482721

ABSTRACT

The unit is capable of providing the information on the man parameters of the radioactive sources, thus improving the reliability and accuracy when planning the treatment in radiotherapy.


Subject(s)
Radiation Monitoring/instrumentation , Radiotherapy/instrumentation , Equipment Design , Humans , Radiotherapy Dosage
20.
Med Tekh ; (3): 40-2, 1977.
Article in Russian | MEDLINE | ID: mdl-407424

ABSTRACT

The betatron B5M25 bremsstrahlung is formed only by a rectangularly shaped diaphragm, which results in excessive irradiation of healthy tissues surrounding the pathological focus. An appliance in the form of an additional diaphragm with inserted shaped units (blocks) fastened to the proper betatron diaphragm and permitting formation of a dose field in conformity with the pathological focus outlines has been constructed. The design of the appliance helps avoid "contamination" of the bremsstrahlung beam with electrons and second generation Compton quanta.


Subject(s)
Radiotherapy, High-Energy/instrumentation , Radiotherapy Dosage , USSR