J Magn Reson ; 334: 107110, 2022 01.
Article in English | MEDLINE | ID: mdl-34844075

ABSTRACT

We present a proof-of-concept study focusing on a method for the intra- and inter-center validation and harmonization of data obtained from MRI T1 and T2 maps. The method is based on a set of MnCl2 samples that provide in-scan ground-truth reference values regardless of the details of the MRI protocol. The relaxation times of MnCl2 aqueous solutions were first measured by means of an NMR laboratory relaxometer, as a function of concentration and temperature. The obtained T1 and T2 values, once renormalized at the scanner temperature, were used as reference values for the MRI mapping measurements of the MnCl2 relaxation times. By using different clinical MRI scanners and sequences, we found a good agreement for standard and turbo sequences (limits of agreement: 5% for IR, SE, IR-TSE; 10% for TSE), while an under-estimation and an over-estimation were found respectively for MOLLI and T2-prep TrueFISP, as already reported in the literature. The linearity of the relaxation rates with the concentration predicted by the Solomon-Bloembergen-Morgan theory was observed for every dataset at all temperatures, except for T2-prep TrueFISP maps results. Some preliminary results of an in vivo experiment are also presented.


Subject(s)
Magnetic Resonance Imaging , Water , Magnetic Resonance Spectroscopy , Reproducibility of Results