Retina ; 31(10): 2002-6, 2011 Nov.
Article in English | MEDLINE | ID: mdl-21673614

ABSTRACT

PURPOSE: To determine and compare 20-, 23-, and 25-gauge retinal infusion air jet impact pressure (force per unit area) in an experimental setting. METHODS: Experimental laboratory investigation. Infusion cannulas were connected to a compressed air system. A controlled valve mechanism was used to obtain increasing levels of infusion pressure. Each infusion tube was positioned in front of a manual transducer to measure force. Impact pressure was calculated using known formulas in fluid dynamics. RESULTS: The 20-gauge infusion jet showed similar impact pressure values compared with the 23-gauge infusion jet. Both showed higher levels than the 25-gauge infusion jet. This was because of the smaller jet force for the 25-gauge system. CONCLUSION: In this experimental study, both the 23- and the 20-gauge air infusion jet showed higher impact pressure values compared with the 25-gauge air infusion jet. This could be of concern regarding air infusion during 23-gauge vitrectomy since retinal damage has been shown in standard-gauge surgeries.


Subject(s)
Catheterization/instrumentation , Compressed Air/adverse effects , Microsurgery/instrumentation , Pressure , Vitrectomy , Humans , Hydrodynamics , Infusions, Parenteral , Models, Theoretical