Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters










Database
Language
Publication year range
1.
Talanta ; 82(1): 417-21, 2010 Jun 30.
Article in English | MEDLINE | ID: mdl-20685487

ABSTRACT

The anodic oxidation of concentrated glycine based aqueous electrolyte on smooth platinum electrode leads to a strongly grafted polyglycine-like coating on the surface in an irreversible way. Due to the proton affinity towards amino groups of polyglycine (PG), the electrodeposited thin film was used as receptor for solid potentiometric pH sensor. In order to reach local pH measurement, we developed miniaturized microelectrodes on glass substrate thanks to photolithography process. We used silver chloride on silver as the reference electrode. The couple (silver chloride, PG based platinum electrode) of microelectrodes gives linear potentiometric response vs. pH in the range [2-12], reversibly and with a sensitivity of 52.4 mV/pH (for 1mm electrode size). PG based pH electrode is compared to other organic polymer based pH receptor such as linear polyethylenimine (L-PEI), polyaniline (PANI) and glass membrane.


Subject(s)
Chemistry Techniques, Analytical/instrumentation , Peptides/chemistry , Platinum/chemistry , Electrochemistry , Electrodes , Hydrogen-Ion Concentration , Oxidation-Reduction , Potentiometry , Quality Control
SELECTION OF CITATIONS
SEARCH DETAIL
...