Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 4 de 4
Filter
Add more filters











Publication year range
1.
Arq Bras Cardiol ; 86(4): 261-7, 2006 Apr.
Article in Portuguese | MEDLINE | ID: mdl-16680290

ABSTRACT

OBJECTIVE: To present a technical proposal based on the experience of 130 implantations using a simplified technique for coronary sinus catheterization, based on the atrial component of the intracavitary electrogram and radiological anatomy. METHODS: From October, 2001 to October, 2004, 130 biventricular pacemaker implantations were performed, using radiological anatomy and observation of the intracavitary electrogram, focusing on the atrial component. RESULTS: The implantation of the system using left ventricular pacing via coronary sinus was not possible in 8 patients. Difficulties on the cannulation of the coronary ostium were felt in 12 patients and difficulties of lead advancement through the coronary sinus were felt in 15 patients. The mean time of radioscopy utilization was 18.69 min. CONCLUSION: The implantation technique, using the atrial component morphology of the intracavitary electrogram and radiological anatomy showed to be workless, safe and effective for the cannulation of the coronary sinus ostium requesting reduced time of radioscopy.


Subject(s)
Cardiac Catheterization/methods , Electrocardiography , Heart Conduction System/physiology , Heart Failure/therapy , Pacemaker, Artificial , Adult , Aged , Aged, 80 and over , Bundle-Branch Block/physiopathology , Bundle-Branch Block/therapy , Cardiomyopathy, Dilated/physiopathology , Cardiomyopathy, Dilated/therapy , Female , Heart Atria/physiopathology , Heart Conduction System/physiopathology , Heart Failure/physiopathology , Heart Ventricles/physiopathology , Humans , Male , Middle Aged , Radiography , Retrospective Studies , Sinoatrial Node/diagnostic imaging , Sinoatrial Node/physiopathology , Treatment Outcome , Ventricular Function, Left/physiology
2.
Arq. bras. cardiol ; Arq. bras. cardiol;86(4): 261-267, abr. 2006. ilus
Article in Portuguese | LILACS | ID: lil-426202

ABSTRACT

OBJETIVO: Apresentar uma proposição técnica baseada na experiência de 130 implantes utilizando técnica simplificada para cateterização do seio coronariano, baseada no componente atrial do eletrograma intracavi-tário e anatomia radiológica. MÉTODOS: De outubro de 2001 a outubro de 2004 foram realiza-dos 130 implantes de marcapasso biventricular, utilizando-se anatomia radiológica e observação de eletrograma intracavitário, com prioridade ao componente atrial. RESULTADOS: O implante do sistema, utilizando-se a estimulação do ventrículo esquerdo via seio coronariano, não foi possível em 8 pacientes. Em 12 pacientes foram observadas dificuldades na canulação do óstio coronário e em 15 pacientes observaram-se dificuldades de progressão do eletrodo através do seio coronariano. O tempo médio de utilização de radioscopia foi de 18,69 min. CONCLUSÃO: A técnica de implante, utilizando a morfologia do componente atrial do eletrograma intracavitário e anatomia radiológica, demonstrou ser pouco trabalhosa, segura e eficaz para canulação do óstio do seio coronariano, necessitando de reduzido tempo de radioscopia.


Subject(s)
Humans , Male , Female , Adult , Middle Aged , Aged, 80 and over , Cardiac Catheterization/methods , Electrocardiography , Heart Failure/therapy , Pacemaker, Artificial , Heart Conduction System/physiology , Heart Atria/physiopathology , Bundle-Branch Block/physiopathology , Bundle-Branch Block/therapy , Cardiomyopathy, Dilated/physiopathology , Cardiomyopathy, Dilated/therapy , Ventricular Function, Left/physiology , Heart Failure/physiopathology , Sinoatrial Node/physiopathology , Sinoatrial Node , Retrospective Studies , Heart Conduction System/physiopathology , Treatment Outcome , Heart Ventricles/physiopathology
3.
Rev. bras. cir. cardiovasc ; Rev. bras. cir. cardiovasc;20(3): 301-309, jul.-set. 2005. ilus
Article in Portuguese | LILACS | ID: lil-421610

ABSTRACT

OBJETIVO: Apresentar a experiência de 157 implantes utilizando uma técnica simplificada para cateterizacão do seio coronariano, baseada no eletrograma intracavitário e anatomia radiológica, demonstrando o porcentual de sucesso e tempo total de utilizacão de radioscopia. MÉTODO: De outubro de 2001 a fevereiro de 2005, foram realizados 157 implantes de marcapasso biventricular em pacientes previamente selecionados, utilizando-se anatomia radiológica e observacão de eletrograma intracavitário, demonstrando a taxa de sucesso, complicacões e tempo total de utilizacão de radioscopia. RESULTADOS: O implante do sistema, utilizando-se a estimulacão do ventrículo esquerdo via seio coronariano, não foi possível em onze procedimentos. Em 20 pacientes foram observadas dificuldades na canulacão do óstio coronário e em 39 pacientes observou-se dificuldade de progressão do eletrodo através do seio coronário. O tempo médio de utilizacão de radioscopia foi 18,27 n 15,46 min. CONCLUSAO: A técnica de implante, proposta pelo autor, utilizando o eletrograma intracavitário e anatomia radiológica, demonstrou ser segura e eficaz para canulacão do óstio do seio coronário, necessitando de reduzidos tempos de radioscopia.


Subject(s)
Humans , Cardiomyopathy, Dilated , Heart Failure/therapy , Cardiac Pacing, Artificial
4.
REBLAMPA Rev. bras. latinoam. marcapasso arritmia ; 16(4): 212-219, out.-dez. 2003. ilus, graf
Article in Portuguese | LILACS | ID: lil-364769

ABSTRACT

Os modelos experimentais permitem o estudo da atuação do marcapasso artificial em fenômenos fisiológicos do coração e sua correlação com a prática clínica. Os objetivos deste trabalho foram padronizar a técnica de obtençao de Bloqueio atrioventricular Total (BAVT)experimental, o estudo do Débito Cardíaco (DC), dos limiares de estimulação, da amplitude de R, da impedância elétrica e a variação da largura do complexo QRS, com o coração em rítmo sinusal em BAVT e estimulação cardíaca epimiocárdica em ponta do ventrículo direito (VD), ventrículo esquerdo (VE) e biventricular, em suínos com o coração supostamente normal. Utilizamos seis animais para obter BAVT experimental pela técnica de esternotomia, sem atriotomia, e necrose química do sistema de condução por injeção de formaldeídeo a 40 por cento. O DC foi medido por termodiluição. O eletrocardiograma foi registrado nas derivações D1, D2 e D3. Os eletrodos epimiocárdicos utilizados foram modelos ELC-35 (Biotronik Gmbh). A estimulação foi feita com equipamento ERA300 (Biotronik) em modo VOO com 5 volts (V) de amplitude e 1 milissegundo (ms) de duração com frequência de estimulação semelhante encontrada quando o coração estava em rítmo sinusal e o tórax aberto...


Subject(s)
Animals , Cardiac Pacing, Artificial , Heart Block , Atropine , Pancuronium/administration & dosage , Pentobarbital , Treatment Outcome
SELECTION OF CITATIONS
SEARCH DETAIL