Anal Chem ; 74(24): 6413-7, 2002 Dec 15.
Article in English | MEDLINE | ID: mdl-12510768

ABSTRACT

A microarray hybridization system that allows mixing in volumes comparable to those used by glass coverslips is presented. This system is composed of a disposable flexible lid that binds to 1 in. x 3 in. glass slides via an adhesive gasket, forming a uniform 25-microm-thick hybridization chamber. This chamber rests on a base unit for temperature control. The lid contains two air-driven bladders that continuously mix the hybridization fluid. Mixing enhances sensitivity from a typical microarray experiment 2-3-fold. Mixing is particularly effective at high spotted probe and low labeled target concentrations and overcoming local target depletion that occurs when homologous probes are spotted in close proximity. Mixing appears to be compatible with most hybridization conditions; however, mix versus no-mix control experiments should be performed. Also covered are a number of microfluidic issues related to manufacturing, filling, mixing, and packaging.


Subject(s)
Nucleic Acid Hybridization/methods , Oligonucleotide Array Sequence Analysis/instrumentation , Oligonucleotide Array Sequence Analysis/methods , Bacteriophage lambda/genetics , Cell Line , DNA/analysis , DNA/chemistry , Genome, Viral , Humans , Polymerase Chain Reaction , Sensitivity and Specificity , Solutions