1.
Radiol Phys Technol ; 17(2): 518-526, 2024 Jun.
Article in English | MEDLINE | ID: mdl-38367143

ABSTRACT

While some MRI systems offer a "pause" function, combining it with the PROPELLER method for image quality improvement remains underexplored. This study investigated whether repositioning the head after pausing during PROPELLER imaging enhances image quality. All brain phantom images in this study were obtained using a 3.0 T MRI and acquired using the fast spin-echo T2WI-based PROPELLER with motion correction. By combining the angle of rotational motion of the head phantom and the number of repositioning after a pause, two studies including seven trials were performed. Increasing the rotation angle decreased the image quality; however, pausing the image and repositioning the head phantom to the original angle improved the image quality. A similar result was obtained by repositioning the angle closer to its original angle. Experiments with multiple head movements showed that pausing the scan and repositioning the phantom with each movement improved image quality.


Subject(s)
Brain , Image Processing, Computer-Assisted , Magnetic Resonance Imaging , Phantoms, Imaging , Magnetic Resonance Imaging/instrumentation , Magnetic Resonance Imaging/methods , Brain/diagnostic imaging , Brain/physiology , Image Processing, Computer-Assisted/methods , Humans , Movement
SELECTION OF CITATIONS