PLoS One ; 11(4): e0152184, 2016.
Article in English | MEDLINE | ID: mdl-27046200

ABSTRACT

In South Africa, various point-of-care hemoglobin meters are used. However, the regulatory framework for approval, implementation and oversight of use of point-of-care hemoglobin meters is suboptimal. We assessed the diagnostic accuracy of the HemoCue Hb 301, STAT-Site MHgb and URIT-12 point-of-care hemoglobin meters, compared to a central laboratory based reference assay, in a central laboratory and a community based clinic in Durban, South Africa. Differences in performance of the point-of-care assays, compared to the reference assay, were more pronounced in the community based clinic. Results were reasonable for the HemoCue Hb 301, but poor for the STAT-Site MHgb and the URIT-12. Poor test performance of point-of-care hemoglobin meters, and inadequate evaluations and oversight in South Africa, leads to suboptimal clinical care and clinical research, and increased costs. There is a need for proper evaluation and quality assurance of point-of-care tests, the results of which should be made widely available to key stakeholders.


Subject(s)
Anemia/blood , Anemia/diagnosis , Hemoglobins/metabolism , Hospitals, Community , Point-of-Care Systems , Adolescent , Adult , Humans , Male , Middle Aged , South Africa/epidemiology