ABSTRACT
14,620 sets of patient dose data were obtained for 31 different models of computed tomography (CT) equipment (total 73) with 18 types of CT examination in Korea. Specific diagnostic reference levels (DRLs) for this study in terms of third quartile volumetric CT dose index in mGy [and dose-length product (DLP) in mGy.cm] are as follows: head, 53 (910); neck, 20 (770); chest, 14 (710); abdomen, 14 (1000); stomach, 14 (1000); liver, 14 (1700); pancreas, 14 (1700); kidney, 14 (2100); cervical spine, 30 (600); lumbar spine, 25 (760); hip, 17 (600); cardiac CT angiography, 45 (1250); head CT angiography, 43 (1900); liver CT angiography, 14 (1400) and thoraco-abdominal CT angiography, 16 (2000). In the present study, DRLs in terms of volumetric CT dose index were below previously published reference levels, partly because the newer CT equipments have improved technology that facilitates lower patient dose. Meanwhile, DRLs in terms of DLP were higher, because multi-phase scanning protocols with prolonged scan coverage have been widely used.