Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters










Database
Language
Publication year range
1.
Pak J Pharm Sci ; 25(1): 135-40, 2012 Jan.
Article in English | MEDLINE | ID: mdl-22186321

ABSTRACT

The objective of this study is to develop sensitive and cost effective reverse phase high performance liquid chromatographic method for the estimation of Metoclopramide Hydrochloride in oral solid dosage formulations. A reverse chromatographic method was used with the mobile phase of Acetonitrile, 20m M Potassium dihydrogen phosphate buffer solution (pH 3 adjusted with orthophosphoric acid) in the ratio of 40:60. The column used was Waters C18 3.9×300mm µBondapak (RP). The flow rate of the mobile phase was 2ml/minute. The detector was set at the wavelength of 275nm.This method showed good sensitivity. The linearity was also found to be excellent (γ(2)=0.997) in the range of 5-75 µg/ml. No interfering peaks were observed at the retention time of Metoclopramide Hydrochloride when both placebo and blank samples were injected (Retention time =1.93min). The parameters such as specificity, linearity, range, accuracy, precision, system suitability, solution stability, detection and quantification limits were evaluated to validate this method. This method can effectively be used for quantitative analysis of Metoclopramide hydrochloride tablet formulations because of its specificity, accuracy and convenience of use.


Subject(s)
Chemistry, Pharmaceutical/methods , Chromatography, High Pressure Liquid/methods , Chromatography, Reverse-Phase/methods , Metoclopramide/analysis , Tablets/chemistry , Chemistry, Pharmaceutical/statistics & numerical data , Drug Stability , Limit of Detection
SELECTION OF CITATIONS
SEARCH DETAIL
...