ABSTRACT

OBJECTIVE: The aim of this report is to propose a scheme for validation of an analytical technique according to ISO 17025. MATERIAL AND METHODS: According to ISO 17025, the fundamental parameters tested were: selectivity, calibration model, precision, accuracy, uncertainty of measurement, and analytical interference. RESULTS: A protocol has been developed that has been applied successfully to quantify zinc in serum by atomic absorption spectrometry. CONCLUSION: It is demonstrated that our method is selective, linear, accurate, and precise, making it suitable for use in routine diagnostics.


Subject(s)
Blood Chemical Analysis/standards , Spectrophotometry, Atomic/methods , Zinc/blood , Accreditation/standards , Calibration , Humans , International Agencies/standards , Observer Variation , Reproducibility of Results , Sensitivity and Specificity , Spectrophotometry, Atomic/instrumentation , Spectrophotometry, Atomic/standards