1.
Anal Chem ; 92(8): 5862-5870, 2020 04 21.
Article in English | MEDLINE | ID: mdl-32212635

ABSTRACT

Mobility isolated spectra were obtained for protonated monomers of 42 volatile oxygen containing organic compounds at ambient pressure using a tandem ion mobility spectrometer with a reactive stage between drift regions. Fragment ions of protonated monomers of alcohols, acetates, aldehydes, ketones, and ethers were produced in the reactive stage using a 3.3 MHz symmetrical sinusoidal waveform with an amplitude of 1.4 kV and mobility analyzed in a 19 mm long drift region. The resultant field induced fragmentation (FIF) spectra included residual intensities for protonated monomers and fragment ions with characteristic drift times and peak intensities, associated with ion mass and chemical class. High efficiency of fragmentation was observed with single bond cleavage of alcohols and in six-member ring rearrangements of acetates. Fragmentation was not observed, or seen weakly, with aldehydes, ethers, and ketones due to their strained four-member ring transition states. Neural networks were trained to categorize spectra by chemical class and tested with FIF spectra of both familiar and unfamiliar compounds. Rates of categorization were class dependent with best performance for alcohols and acetates, moderate performance for ketones, and worst performance for ethers and aldehydes. Trends in the rates of categorization within a chemical family can be understood as steric influences on the energy of activation for ion fragmentation. Electric fields greater than 129 Td or new designs of reactive stages with improved efficiency of fragmentation will be needed to extend the practice of reactive stage tandem IMS to an expanded selection of volatile organic compounds.

2.
J Am Soc Mass Spectrom ; 28(10): 2151-2159, 2017 10.
Article in English | MEDLINE | ID: mdl-28664477

ABSTRACT

We provide modeling and experimental data describing the dominant ion-loss mechanisms for differential mobility spectrometry (DMS). Ion motion is considered from the inlet region of the mobility analyzer to the DMS exit, and losses resulting from diffusion to electrode surfaces, insufficient effective gap, ion fragmentation, and fringing field effects are considered for a commercial DMS system with 1-mm gap height. It is shown that losses due to diffusion and radial oscillations can be minimized with careful consideration of residence time, electrode spacing, gas flow rate, and waveform frequency. Fragmentation effects can be minimized by limitation of the separation field. When these parameters were optimized, fringing field effects at the DMS inlet contributed the most to signal reduction. We also describe a new DMS cell configuration that improves the gas dynamics at the mobility cell inlet. The new cell provides a gas jet that decreases the residence time for ions within the fringing field region, resulting in at least twofold increase in ion signal as determined by experimental data and simulations. Graphical Abstract ᅟ.

3.
Anal Chem ; 88(23): 11399-11405, 2016 12 06.
Article in English | MEDLINE | ID: mdl-27715016

ABSTRACT

Differential mobility spectrometry (DMS) is applied to the analysis of inorganic mixtures relevant to nuclear forensics. Three primary components of potential radiological dispersal devices (RDDs), cobalt, cesium, and strontium, were studied by DMS to demonstrate rapid sample cleanup when coupled to mass spectrometry. Nanosprayed salt solutions comprised of stable analogs, as proxies to these radioisotopes, and isobaric interferents were introduced to DMS. The DMS effluent was directly coupled to a mass spectrometer to confirm the elemental identity of the separated clusters. DMS dispersion plots demonstrated distinctive elemental separation from both atomic and molecular interferents. These results support the potential use of DMS as a means of rapid separation for inorganic analyses prior to analysis in a field portable mass spectrometer. The mechanism for this process is speculated to involve dynamics of solvent cluster formation under the influence of the alternating high and low electric fields of the DMS.

4.
Mass Spectrom Rev ; 35(6): 687-737, 2016 10.
Article in English | MEDLINE | ID: mdl-25962527

ABSTRACT

This review of differential mobility spectrometry focuses primarily on mass spectrometry coupling, starting with the history of the development of this technique in the Soviet Union. Fundamental principles of the separation process are covered, in addition to efforts related to design optimization and advancements in computer simulations. The flexibility of differential mobility spectrometry design features is explored in detail, particularly with regards to separation capability, speed, and ion transmission. 2015 Wiley Periodicals, Inc. Mass Spec Rev 35:687-737, 2016.


Subject(s)
Mass Spectrometry , Computer Simulation , Spectrum Analysis
5.
Int J Ion Mobil Spectrom ; 16(3): 217-227, 2013 Sep.
Article in English | MEDLINE | ID: mdl-23914140

ABSTRACT

Electrospray ionization (ESI) is an important tool in chemical and biochemical survey and targeted analysis in many applications. For chemical detection and identification electrospray is usually used with mass spectrometry (MS). However, for screening and monitoring of chemicals of interest in light, low power field-deployable instrumentation, an alternative detection technology with chemical selectivity would be highly useful, especially since small, lightweight, chip-based gas and liquid chromatographic technologies are being developed. Our initial list of applications requiring portable instruments includes chemical surveys on Mars, medical diagnostics based on metabolites in biological samples, and water quality analysis. In this report, we evaluate ESI-Differential Mobility Spectrometry (DMS) as a compact, low-power alternative to MS detection. Use of DMS for chemically-selective detection of ESI suffers in comparison with mass spectrometry because portable MS peak capacity is greater than that of DMS by 10X or more, but the development of light, fast chip chromatography offers compensating resolution. Standalone DMS provides the chemical selectivity familiar from DMS-MS publications, and exploits the sensitivity of ion detection. We find that sub-microliter-per-minute flows and a correctly-designed interface prepare a desolvated ion stream that enables DMS to act as an effective ion filter. Results for a several small organic biomarkers and metabolites, including citric acid, azelaic acid, n-hexanoylglycine, thymidine, and caffeine, as well as compounds such as dinitrotoluene and others, have been characterized and demonstrate selective detection. Water-quality-related halogen-containing anions, fluoride through bromate, contained in liquid samples are also isolated by DMS. A reaction-chamber interface is highlighted as most practical for portable ESI-DMS instrumentation.

6.
J Am Soc Mass Spectrom ; 24(9): 1428-36, 2013 Sep.
Article in English | MEDLINE | ID: mdl-23797861

ABSTRACT

A miniature, planar, differential ion mobility spectrometer (DMS) was interfaced to an LCQ classic ion trap to conduct selective ion filtration prior to mass analysis in order to extend the dynamic range of the trap. Space charge effects are known to limit the functional ion storage capacity of ion trap mass analyzers and this, in turn, can affect the quality of the mass spectral data generated. This problem is further exacerbated in the analysis of mixtures where the indiscriminate introduction of matrix ions results in premature trap saturation with non-targeted species, thereby reducing the number of parent ions that may be used to conduct MS/MS experiments for quantitation or other diagnostic studies. We show that conducting differential mobility-based separations prior to mass analysis allows the isolation of targeted analytes from electrosprayed mixtures preventing the indiscriminate introduction of matrix ions and premature trap saturation with analytically unrelated species. Coupling these two analytical techniques is shown to enhance the detection of a targeted drug metabolite from a biological matrix. In its capacity as a selective ion filter, the DMS can improve the analytical performance of analyzers such as quadrupole (3D or linear) and ion cyclotron resonance (FT-ICR) ion traps that depend on ion accumulation.


Subject(s)
Tandem Mass Spectrometry/instrumentation , Cocaine/analogs & derivatives , Cocaine/urine , Equipment Design , Humans , Ions/chemistry
7.
J Forensic Sci ; 57(3): 750-6, 2012 May.
Article in English | MEDLINE | ID: mdl-22235847

ABSTRACT

Forensic drug laboratories are inundated with cases requiring time-consuming GC- or LC-based chromatographic separations of submitted samples. High-throughput analytical methods would be of great practical utility within forensic drug analysis. Recently developed ion-mobility-based separation methods combined with mass spectrometry can often be used without chromatography, suppress chemical interferents of similar mass, and operate in seconds. We have evaluated differential mobility spectrometry-mass spectrometry (DMS-MS) for performance on adulterated cocaine mixtures. The DMS interface is only a few centimeters in length, operates in seconds, and can be adapted to any MS system using atmospheric pressure ionization. Drug cutting agents, typical targets such as cocaine, and drug metabolites are rapidly separated by the DMS ion prefilter. Tests demonstrated characterization of complex mixtures, such as isolation of levamisole, an adulterant with alarming side effects, from a 13-component mixture. DMS-MS holds great potential for the analysis of drug samples submitted for forensic analysis.

8.
Int J Ion Mobil Spectrom ; 15(3)2012 Sep 01.
Article in English | MEDLINE | ID: mdl-24311968

ABSTRACT

Clinical and forensic toxicology laboratories are inundated with thousands of samples requiring lengthy chromatographic separations prior to mass spectrometry. Here, we employ differential mobility spectrometry (DMS) interfaced to nano-electrospray ionization-mass spectrometry to provide a rapid ion filtration technique for the separation of ions in gas phase media prior to mass spectral analysis on a DMS-integrated AB SCIEX API 3000 triple-quadrupole mass spectrometer. DMS is efficient at the rapid separation of ions under ambient conditions and provides many advantages when used as an ion filtration technique in tandem with mass spectrometry (MS) and MS/MS. Our studies evaluated DMS-MS/MS as a rapid, quantitative platform for the analysis of drug metabolites isolated from urine samples. In targeted applications, five metabolites of common drugs of abuse were effectively and rapidly separated using isopropanol and ethyl acetate as transport gas modifiers, eliminating the gas chromatography or liquid chromatography-based separations commonly employed in clinical and forensic toxicology laboratories. Calibration curves were prepared for the selected drug metabolites utilizing deuterated internal standards for quantitative purposes. The feasibility of separating and quantitating drug metabolites in a rapid fashion was evaluated by compensation voltage stepping followed by multiple reaction monitoring (MRM) detection. Rapid profiling of clinical and forensic toxicology samples could help to address an urgent need within the scientific community by developing high-throughput analytical methodologies, which could reduce significant case backlogs present within these laboratories.

9.
J Am Soc Mass Spectrom ; 21(9): 1477-81, 2010 Sep.
Article in English | MEDLINE | ID: mdl-20619673

ABSTRACT

A planar differential mobility spectrometer (DMS) was coupled to a Mini 10 handheld rectilinear ion trap (RIT) mass spectrometer (MS) (total weight 10 kg), and the performance of the instrument was evaluated using illicit drug analysis. Coupling of DMS (which requires a continuous flow of drift gas) with a miniature MS (which operates best using sample introduction via a discontinuous atmospheric pressure interface, DAPI), was achieved with auxiliary pumping using a 5 L/min miniature diaphragm sample pump placed between the two devices. On-line ion mobility filtering showed to be advantageous in reducing the background chemical noise in the analysis of the psychotropic drug diazepam in urine using nanoelectrospray ionization. The combination of a miniature mass spectrometer with simple and rapid gas-phase ion separation by DMS allowed the characteristic fragmentation pattern of diazepam to be distinguished in a simple urine extract at lower limits of detection (50 ng/mL) than that achieved without DMS (200 ng/mL). The additional separation power of DMS facilitated the identification of two drugs of similar molecular weight, morphine (average MW = 285.34) and diazepam (average MW = 284.70), using a miniature mass spectrometer capable of unit resolution. The similarity in the proton affinities of these two compounds resulted in some cross-interference in the MS data due to facile ionization of the neutral form of the compound even when the ionic form had been separated by DMS.


Subject(s)
Forensic Toxicology/methods , Illicit Drugs/analysis , Spectrometry, Mass, Electrospray Ionization/methods , Substance Abuse Detection/methods , Diazepam/urine , Forensic Toxicology/instrumentation , Humans , Morphine/urine , Nanotechnology , Spectrometry, Mass, Electrospray Ionization/instrumentation , Substance Abuse Detection/instrumentation
10.
Int J Mass Spectrom ; 291(3): 108-117, 2010 Apr 15.
Article in English | MEDLINE | ID: mdl-20305793

ABSTRACT

Technology to enable rapid screening for radiation exposure has been identified as an important need, and, as a part of a NIH / NIAD effort in this direction, metabolomic biomarkers for radiation exposure have been identified in a recent series of papers. To reduce the time necessary to detect and measure these biomarkers, differential mobility spectrometry - mass spectrometry (DMS-MS) systems have been developed and tested. Differential mobility ion filters preselect specific ions and also suppress chemical noise created in typical atmospheric-pressure ionization sources (ESI, MALDI, and others). Differential-mobility-based ion selection is based on the field dependence of ion mobility, which, in turn, depends on ion characteristics that include conformation, charge distribution, molecular polarizability, and other properties, and on the transport gas composition which can be modified to enhance resolution. DMS-MS is able to resolve small-molecule biomarkers from nearly-isobaric interferences, and suppresses chemical noise generated in the ion source and in the mass spectrometer, improving selectivity and quantitative accuracy. Our planar DMS design is rapid, operating in a few milliseconds, and analyzes ions before fragmentation. Depending on MS inlet conditions, DMS-selected ions can be dissociated in the MS inlet expansion, before mass analysis, providing a capability similar to MS/MS with simpler instrumentation. This report presents selected DMS-MS experimental results, including resolution of complex test mixtures of isobaric compounds, separation of charge states, separation of isobaric biomarkers (citrate and isocitrate), and separation of nearly-isobaric biomarker anions in direct analysis of a bio-fluid sample from the radiation-treated group of a mouse-model study. These uses of DMS combined with moderate resolution MS instrumentation indicate the feasibility of field-deployable instrumentation for biomarker evaluation.

11.
Rev Sci Instrum ; 81(2): 024101, 2010 Feb.
Article in English | MEDLINE | ID: mdl-20192506

ABSTRACT

Devices based on differential mobility spectrometry (DMS) are used in a number of ways, including applications as ion prefilters for API-MS systems, as detectors or selectors in hybrid instruments (GC-DMS, DMS-IMS), and in standalone systems for chemical detection and identification. DMS ion separation is based on the relative difference between high field and low field ion mobility known as the alpha dependence, and requires the application of an intense asymmetric electric field known as the DMS separation field, typically in the megahertz frequency range. DMS performance depends on the waveform and on the magnitude of this separation field. In this paper, we analyze the relationship between separation waveform and DMS resolution and consider feasible separation field generators. We examine ideal and practical DMS separation field waveforms and discuss separation field generator circuit types and their implementations. To facilitate optimization of the generator designs, we present a set of relations that connect ion alpha dependence to DMS separation fields. Using these relationships we evaluate the DMS separation power of common generator types as a function of their waveform parameters. Optimal waveforms for the major types of DMS separation generators are determined for ions with various alpha dependences. These calculations are validated by comparison with experimental data.


Subject(s)
Motion , Spectrum Analysis/methods , Electricity , Feasibility Studies , Models, Theoretical , Reproducibility of Results , Spectrum Analysis/instrumentation
12.
Anal Chem ; 82(5): 1867-80, 2010 Mar 01.
Article in English | MEDLINE | ID: mdl-20121077

ABSTRACT

In differential mobility spectrometry (also referred to as high-field asymmetric waveform ion mobility spectrometry), ions are separated on the basis of the difference in their mobility under high and low electric fields. The addition of polar modifiers to the gas transporting the ions through a differential mobility spectrometer enhances the formation of clusters in a field-dependent way and thus amplifies the high- and low-field mobility difference, resulting in increased peak capacity and separation power. Observations of the increase in mobility field dependence are consistent with a cluster formation model, also referred to as the dynamic cluster-decluster model. The uniqueness of chemical interactions that occur between an ion and cluster-forming neutrals increases the selectivity of the separation, and the depression of low-field mobility relative to high-field mobility increases the compensation voltage and peak capacity. The effect of a polar modifier on the peak capacity across a broad range of chemicals has been investigated. We discuss the theoretical underpinnings which explain the observed effects. In contrast to the result with a polar modifier, we find that using mixtures of inert gases as the transport gas improves the resolution by reducing the peak width but has very little effect on the peak capacity or selectivity. The inert gas helium does not cluster and thus does not reduce low-field mobility relative to high-field mobility. The observed changes in the differential mobility alpha parameter exhibited by different classes of compounds when the transport gas contains a polar modifier or has a significant fraction of inert gas can be explained on the basis of the physical mechanisms involved in the separation processes.


Subject(s)
Mass Spectrometry/instrumentation
13.
Article in English | MEDLINE | ID: mdl-20065515

ABSTRACT

Differential mobility spectrometry (DMS) separates ions on the basis of the difference in their migration rates under high versus low electric fields. Several models describing the physical nature of this field mobility dependence have been proposed but emerging as a dominant effect is the clusterization model sometimes referred to as the dynamic cluster-decluster model. DMS resolution and peak capacity is strongly influenced by the addition of modifiers which results in the formation and dissociation of clusters. This process increases selectivity due to the unique chemical interactions that occur between an ion and neutral gas-phase molecules. It is thus imperative to bring the parameters influencing the chemical interactions under control and find ways to exploit them in order to improve the analytical utility of the device. In this paper, we describe three important areas that need consideration in order to stabilize and capitalize on the chemical processes that dominate a DMS separation. The first involves means of controlling the dynamic equilibrium of the clustering reactions with high concentrations of specific reagents. The second area involves a means to deal with the unwanted heterogeneous cluster ion populations emitted from the electrospray ionization process that degrade resolution and sensitivity. The third involves fine control of parameters that affect the fundamental collision processes, temperature and pressure.


Subject(s)
Mass Spectrometry/instrumentation , Mass Spectrometry/methods , Ions/chemistry , Pressure , Spectrometry, Mass, Electrospray Ionization/instrumentation , Spectrometry, Mass, Electrospray Ionization/methods , Temperature
14.
Int J Mass Spectrom ; 298(1-3): 45-54, 2010 Dec 01.
Article in English | MEDLINE | ID: mdl-21278836

ABSTRACT

Ion filters based on planar DMS can be integrated with the inlet configuration of most mass spectrometers, and are able to enhance the quality of mass analysis and quantitative accuracy by reducing chemical noise, and by pre-separating ions of similar mass. This paper is the first in a series of three papers describing the optimization of DMS / MS instrumentation. In this paper the important physical parameters of a planar DMS-MS interface including analyzer geometry, analyzer coupling to a mass spectrometer, and transport gas flow control are considered. The goal is to optimize ion transmission and transport efficiency, provide optimal and adjustable resolution, and produce stable operation under conditions of high sample contamination. We discuss the principles of DMS separations and highlight the theoretical underpinnings. The main differences between planar and cylindrical geometries are presented, including a discussion of the advantages and disadvantages of RF ion focusing. In addition, we present a description of optimization of the frequency and amplitude of the DMS fields for resolution and ion transmission, and a discussion of the influence and importance of ion residence time in DMS. We have constructed a mass spectrometer interface for planar geometries that takes advantage of atmospheric pressure gas dynamic principles, rather than ion focusing, to minimize ion losses from diffusion in the analyzer and to maximize total ion transport into the mass spectrometer. A variety of experimental results has been obtained that illustrate the performance of this type of interface, including tests of resistance to high contamination levels, and the separation of stereoisomers. In a subsequent publication the control of the chemical interactions that drive the separation process of a DMS / MS system will be considered. In a third publication we describe novel electronics designed to provide the high voltages asymmetric waveform fields (SV) required for these devices as well as the effects of different waveforms.

15.
J Am Soc Mass Spectrom ; 18(3): 502-11, 2007 Mar.
Article in English | MEDLINE | ID: mdl-17141523

ABSTRACT

Differential mobility spectrometry (DMS), also commonly referred to as high field asymmetric waveform ion mobility spectrometry (FAIMS) is a rapidly advancing technology for gas-phase ion separation. The interfacing of DMS with mass spectrometry (MS) offers potential advantages over the use of mass spectrometry alone. Such advantages include improvements to mass spectral signal/noise, orthogonal/complementary ion separation to mass spectrometry, enhanced ion and complexation structural analysis, and the potential for rapid analyte quantitation. In this report, we demonstrate the successful use of our nanoESI-DMS-MS system, with a methanol drift gas modifier, for the separation of oligosaccharides. The tendency for ESI to form oligosaccharide aggregate ions and the negative impact this has on nanoESI-DMS-MS oligosaccharide analysis is described. In addition, we demonstrate the importance of sample solvent selection for controlling nanoESI oligosaccharide aggregate ion formation and its effect on glycan ionization and DMS separation. The successful use of a tetrachloroethane/methanol solvent solution to reduce ESI oligosaccharide aggregate ion formation while efficiently forming a dominant MH(+) molecular ion is presented. By reducing aggregate ion formation in favor of a dominant MH(+) ion, DMS selectivity and specificity is improved. In addition to DMS, we would expect the reduction in aggregate ion complexity to be beneficial to the analysis of oligosaccharides for other post-ESI separation techniques such as mass spectrometry and ion mobility. The solvent selected control over MH(+) molecular ion formation, offered by the use of the tetrachloroethane/methanol solvent, also holds promise for enhancing MS/MS structural characterization analysis of glycans.


Subject(s)
Oligosaccharides/analysis , Solvents/chemistry , Spectrometry, Mass, Electrospray Ionization , Mass Spectrometry , Oligosaccharides/chemistry
16.
Anal Chem ; 78(22): 7697-706, 2006 Nov 15.
Article in English | MEDLINE | ID: mdl-17105161

ABSTRACT

A microfabricated planar differential ion mobility spectrometer operating from 0.4 to 1.55 atm in a supporting atmosphere of purified air was used to characterize the effects of pressure and electric field strength on compensation voltage, ion transmission, peak width, and peak intensity in differential mobility spectra. Peak positions, in compensation voltage as a function of separating rf voltage, were found to vary with pressure in a way that can be simplified by expressing both compensation and separation fields in Townsend units for E/N. The separation voltage providing the greatest compensation voltage and the greatest resolution is ion-specific but often occurs at E/N values that are unreachable at elevated pressure because of electrical breakdown. The pressure dependence of air breakdown voltage near 1 atm is sublinear, allowing higher E/N values to be reached at reduced pressure, usually resulting in greater instrumental resolution. Lower voltage requirements at reduced pressure also reduce device power consumption.


Subject(s)
Algorithms , Mass Spectrometry/methods , Microchemistry/methods , Air Pressure , Ions/analysis , Mass Spectrometry/instrumentation , Microchemistry/instrumentation , Organophosphonates/analysis , Organophosphorus Compounds/analysis , Protons , Salicylates/analysis , Sensitivity and Specificity
17.
Anal Chem ; 78(15): 5443-52, 2006 Aug 01.
Article in English | MEDLINE | ID: mdl-16878881

ABSTRACT

Differential mobility spectrometry (DMS) (see Buryakov, I. A.; Krylov, E. V.; Nazarov, E. G.; Rasulev, U. Kh. Int. J. Mass Spectrom. Ion Processes 1993, 128, 143-148), also commonly referred to as high-field asymmetric waveform ion mobility spectrometry (FAIMS) (see Purves, R. W.; Guevremont, R.; Day, S.; Pipich, C. W.; Matyjaszcyk, M. S. Rev. Sci. Instrum. 1998, 69, 4094-4105), is a rapidly advancing technology for gas-phase ion separation. The interfacing of DMS with mass spectrometry (MS) offers potential advantages over the use of mass spectrometry alone. Such advantages include improvements to mass spectral signal-to-noise, orthogonal/complementary ion separation to mass spectrometry, enhanced ion and complexation structural analysis, and the potential for rapid analyte quantitation. In this report, we investigate the use of our nanoESI-DMS-MS system to demonstrate differential mobility separation of peptides. The formation of higher order peptide aggregate ions (ion complexes) via electrospray ionization and the negative impact this has on DMS peptide separation are examined. The successful use of differential mobility drift gas modifiers (dopants) to reduce aggregate ion size and improve DMS peptide ion separation is presented. Following optimization of DMS peptide separation conditions, we examined next the feasibility of a new analytical platform which uses direct sample infusion with nanoESI-DMS-MS for ultrarapid analyte quantitation. Quantitation of a selected peptide from a semicomplex peptide mixture is presented. Initial feasibility results with this new approach demonstrate good accuracy and reproducibility, as well as an absolute mass sensitivity of 6.8 amol and a minimum dynamic range of 2500 for the peptide of interest. This report offers a first look at utilizing nanoESI-DMS-MS to create an ultrarapid (under 5 s) quantitative analysis platform and its potential in the high-throughput arena. Each ion separation technique, DMS and MS, offers orthogonal ion separation to one another, enhancing the overall specificity for this quantitative approach.


Subject(s)
Mass Spectrometry/instrumentation , Mass Spectrometry/methods , Nanotechnology/instrumentation , Nanotechnology/methods , Peptides/analysis , Sensitivity and Specificity
18.
Anal Chem ; 78(13): 4553-63, 2006 Jul 01.
Article in English | MEDLINE | ID: mdl-16808465

ABSTRACT

Positive and negative ion spectra have been obtained with a miniature differential mobility spectrometer equipped with a photoionization source operating at atmospheric pressure. With benzene as a dopant, providing C6H6+ as reactant ion, protonated molecular ions and proton-bound dimer ions were obtained with dimethyl methylphosphonate and butanone. The spectra obtained from gas chromatographic injections of aromatic hydrocarbons, benzene, toluene, and the xylenes, produced the molecular ions when the moisture level was very low, but at a high level the hydrated proton was also present. Possible mechanisms for the formation of protonated products are discussed. Negative ions were produced from electron capture by sulfur hexafluoride using benzene or acetone as dopant. Photoionization of nitrogen dioxide led to the formation of the nitrate ion whose yield was a nonlinear function of concentration. The use of a suitable dopant enhanced ion formation by up to 2 orders of magnitude, and limits of detection in both the positive and negative modes were all at the sub ppm(v) level. The study makes a strong case for the use of a photoionization source as an alternative to the radioactive 63Ni source.

19.
Anal Chem ; 78(1): 96-106, 2006 Jan 01.
Article in English | MEDLINE | ID: mdl-16383315

ABSTRACT

Differential mobility spectrometry (DMS) is a rapidly advancing technology for gas-phase ion separation. The interfacing of DMS with mass spectrometry (MS) offers potential advantages over the use of mass spectrometry alone. Such advantages include improvements to mass spectral signal/noise ratios, orthogonal/complementary ion separation to mass spectrometry, enhanced ion and complexation structural analysis, and potential for rapid analyte quantitation. The introduction of a new ESI-DMS-MS system and its utilization to aid in the understanding of DMS separation theory is described. A current contribution to DMS separation theory is one of an association/dissociation process between ions/molecules in the gas phase during the differential mobility separation. A model study was designed to investigate the molecular dynamics and chemical factors influencing the theorized association/dissociation process, and the mechanisms by which these gas-phase interactions affect an ion's DM behavior. Five piperidine analogues were selected as model analytes, and three alcohol drift gas dopants/modifiers were used to interrogate the analyte ions in the gas phase. Two proposed DMS separation mechanisms, introduced as Core and Façade, corresponding to strong and weak attractions between ions/molecules in the gas phase, are detailed. The proposed mechanisms provide explanation for the observed changes in analyte separation by the various drift gas modifiers. Molecular modeling of the proposed mechanisms provides supportive data and demonstrates the potential for predictive optimization of analyte separation based on drift gas modifier effects.

20.
Anal Chem ; 77(23): 7563-71, 2005 Dec 01.
Article in English | MEDLINE | ID: mdl-16316163

ABSTRACT

A 3.0-m-long, 150-microm-wide, 240-microm-deep channel etched in a 3.2-cm-square silicon chip, covered with a Pyrex wafer, and coated with a dimethyl polysiloxane stationary phase is used for the GC separation of volatile organic compounds. The column, which generates approximately 5500 theoretical plates, is temperature-programmed in a conventional convection oven. The column is connected through a heated transfer line to a microfabricated differential mobility spectrometer. The spectrometer incorporates a 63Ni source for atmospheric-pressure chemical ionization of the analytes. Nitrogen or air transport gas (flow 300 cm(3)/min) drives the analyte ions through the cell. The spectrometer operates with an asymmetric radio frequency (RF) electric field between a pair of electrodes in the detector cell. During each radio frequency cycle, the ion mobility alternates between a high-field and a low-field value (differential mobility). Ions oscillate between the electrodes, and only ions with an appropriate differential mobility reach a pair of biased collectors at the downstream end of the cell. A compensation voltage applied to one of the RF electrodes is scanned to allow ions with different differential mobilities to pass through the cell without being annihilated at the RF electrodes. A unique feature of the device is that both positive and negative ions are detected from a single experiment. The combined microfabricated column and detector is evaluated for the analysis of volatile organic compounds with a variety of functionalities.


Subject(s)
Chromatography, Gas/instrumentation , Chromatography, Gas/methods , Organic Chemicals/chemistry , Silicon/chemistry , Temperature , Volatilization