Vestn Rentgenol Radiol ; (5): 8-11, 1997.
Article in Russian | MEDLINE | ID: mdl-9412052

ABSTRACT

Signs of microdislocation of an endocardial electrode (EE) were observed in 56 (14.4%) of 390 patients with an implantable pacemaker in the postoperative period. The microdislocation was diagnosed from electrocardiographic evidence for no response of a pacing pulse and from fluctuations in the pacing threshold, and from X-ray findings of the thickened and ill-defined outlines of a distal EE with preserving cardiac stimulation. The microdislocation preserves high parameters, as evidenced by a PM-20 M digital analyzer.


Subject(s)
Endocardium/diagnostic imaging , Foreign-Body Migration/diagnostic imaging , Pacemaker, Artificial/adverse effects , Radiography, Thoracic , Adult , Aged , Aged, 80 and over , Arrhythmias, Cardiac/therapy , Electrocardiography , Electrodes , Female , Foreign-Body Migration/etiology , Foreign-Body Migration/physiopathology , Humans , Male , Middle Aged