Int J Radiat Oncol Biol Phys ; 41(3): 721-7, 1998 Jun 01.
Article in English | MEDLINE | ID: mdl-9635725

ABSTRACT

PURPOSE: To assess the accuracy of a conventional simulation procedure in radiotherapy of age-related macular degeneration. METHODS AND MATERIALS: A computed tomographic (CT) extension attached to the treatment simulator was used to acquire CT images immediately after conventional simulation in 18 patients referred for treatment of age-related macular degeneration. Analysis was performed on 16 one-sided treatment cases for whom images were obtained. Error was estimated by the displacement between the observed treatment isocenter and the intended isocenter based on reconstructed eye geometry. RESULTS: Based on single slice measurements, the mean error amplitude was 2.3 mm (range 0.2-5.6). Based on three-dimensional eye globe reconstruction, the mean error amplitude was 2.8 mm (range 0.8-5.3). An incidental finding previously unreported was the lower image quality at the center of the simulator-CT image acquisition field. CONCLUSIONS: Small but significant errors from conventional simulation were noted. The integrated simulation-CT procedure may help correct the errors to improve the accuracy of simulation setup. The lower image quality at the center of image acquisition field requires adaptation of the simulation-CT procedure.


Subject(s)
Macular Degeneration/radiotherapy , Radiotherapy, Computer-Assisted/methods , Tomography, X-Ray Computed , Humans , Macular Degeneration/diagnostic imaging