J Chromatogr A ; 691(1-2): 239-46, 1995 Feb 03.
Article in English | MEDLINE | ID: mdl-7894649

ABSTRACT

Cortisol, cortisone, corticosterone, prednisone and prednisolone are extracted from serum using the novel particle-loaded octyl (C8)-bonded silica in PTFE membrane. Extracts are directly injected, without further concentration, onto a narrow (2.0 mm) or conventional (4.6 mm) bore octyldecyl (C18) HPLC column. Method performance data demonstrate linearity from 0.4 microgram/dl (low limit of detection) up to at least 60 micrograms/dl. Extraction recoveries exceeded 85% and precision (between-run) R.S.D.s averaged < 5%. Interferences were minimal and selectivity was improved over conventional immunochemical steroid assays. When compared to large particle sorbents packed in columns or to traditional liquid-liquid extractions, the membrane extracted steroids in less time, used less reagent, and had smaller elution volumes, thereby obviating steroid instability/adsorption problems associated with traditional concentrating techniques required to improve analytical sensitivity.


Subject(s)
Chromatography, High Pressure Liquid/methods , Membranes, Artificial , Chromatography, High Pressure Liquid/statistics & numerical data , Corticosterone/blood , Cortisone/blood , Drug Stability , Humans , Hydrocortisone/blood , Prednisolone/blood , Prednisone/blood , Sensitivity and Specificity