ABSTRACT
A highly sensitive HPLC method for enantioselective determination of carvedilol in human whole blood and plasma was developed. Carvedilol and S-carazolol as an internal standard extracted from whole blood or plasma were separated using an enantioselective separation column (Chiralpak AD column; 2.0 diameter x 250 mm) without any chiral derivatizations. The mobile phase was hexane:isopropanol:diethylamine (78:22:1, v/v). The excitation and emission wavelengths were set at 284 and 343 nm, respectively. The limits of quantification for the S(-)- and R(+)-carvedilol enantiomers in plasma and blood were both 0.5 ng/ml. Intra- and inter-day variations were less than 5.9%. As an application of the assay, concentrations of carvedilol enantiomer in plasma and blood samples from 15 patients treated with carvedilol for congestive heart failure were determined.