ABSTRACT
The quality of recorded peripheral nerve signals is decisive for their application in therapies. The electroneurogram can be recorded via implantable circumeural electrodes that are wrapped around the peripheral nerve. The shape and amplitude of the signal recorded by the electrode are influenced by the design and contact configuration of the electrode. In this paper, the impact of the number of contacts, contact size and electrical insulation to the outside is investigated to predict the single fiber action potential based on the measured impedance data.