Prikl Biokhim Mikrobiol ; 49(2): 209-14, 2013.
Article in Russian | MEDLINE | ID: mdl-23795482

ABSTRACT

Glucose potentiometric biosensor was prepared by immobilizing glucose oxidase on iodide-selective electrode. The hydrogen peroxide formed after the oxidation of glucose catalysed by glucose oxidase (GOD) was oxidized by sodium molybdate (SMo) at iodide electrode in the presence ofdichlorometane. The glucose concentration was calculated from the decrease of iodide concentration determined by iodide-selective sensor. The sensitivity of glucose biosensor towards iodide ions and glucose was in the concentration ranges of 1.0 x 10(-1) - 1.0 x 10(-6) M and 1.0 x 10(-2) - 1.0 x 10(-4) M, respectively. The characterization of proposed glucose biosensor and glucose assay in human serum were also investigated.


Subject(s)
Biosensing Techniques , Blood Glucose/analysis , Enzymes, Immobilized/chemistry , Glucose Oxidase/chemistry , Iodides/chemistry , Catalysis , Electrodes , Humans , Hydrogen Peroxide/chemistry , Methylene Chloride/chemistry , Molybdenum/chemistry , Oxidation-Reduction , Potentiometry , Reproducibility of Results , Sensitivity and Specificity