Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters










Database
Language
Publication year range
1.
Int J Oral Maxillofac Surg ; 23(6 Pt 2): 418-9, 1994 Dec.
Article in English | MEDLINE | ID: mdl-7890987

ABSTRACT

To investigate titanium surface abrasion of dental implants by mere insertion into the bone, we tested 20 implants of different shape, surface, and length. Under in vitro conditions, it could be proved that longer implants had more abrasion than cylindric implants, that screw implants had more abrasion than cylindric implants, that carved threads caused less abrasion than uncarved ones, and that plasma-flame-coated implants had more abrasion than implants with polished surfaces. Possible causes of the metal abrasion are discussed.


Subject(s)
Dental Implants , Dental Prosthesis Design , Animals , Corrosion , Dental Implantation, Endosseous , Foreign Bodies , Lung , Particle Size , Peritoneum , Ribs , Spleen , Surface Properties , Swine , Titanium/chemistry
SELECTION OF CITATIONS
SEARCH DETAIL
...