1.
Clin Lab ; 55(5-6): 201-6, 2009.
Article in English | MEDLINE | ID: mdl-19728553

ABSTRACT

BACKGROUND: Because of the vast range of physiological relevant estradiol concentrations the requirements to be met by an estradiol assay are high. In the present study the performance of various commercially available estradiol assays was evaluated with regard to imprecision and long-term stability. METHODS: Precision and long-term stability of 7 commercially available estradiol immunoassays were assessed in a multi-centre quality control study based on the repeated measurement of liquid BIOREF estradiol control sera by 18 laboratories during a 14-month study period. RESULTS: The mean estradiol concentrations determined in 594 runs performed for each control level were 71 pg/ml, 349 pg/ml and 676 pg/ml. A high variation was found for the method specific mean values calculated from all results measured with the same method, which ranged between 32 - 90 pg/ml, 187 - 392 pg/ml and 373 - 790 pg/ml, resulting in a similar high inter-laboratory variation with coefficients of variation (CVs) of 25.0%, 16.7% and 17.5%. In contrast, the intra-laboratory variation of estradiol values as well as the variation of values measured with the same method were found to be considerably lower with coefficients of variation < 10% for most laboratories and methods; only the low control level was measured with CV values > 10% by the majority of laboratories and methods. For none of the laboratories a tendency was observed in the results from beginning to end of the 14 month study period indicating a high uniformity in assay production and a good long-term stability of the control material used. CONCLUSIONS: The present data demonstrate that also with the currently available estradiol immunoassays the comparability of results measured with different methods is limited. With most assays very low estradiol concentrations, as observed in postmenopausal women, can be determined only with a precision which is not adequate for clinical assessment.


Subject(s)
Estradiol/blood , Immunoassay/standards , Drug Stability , Female , Follicular Phase/physiology , Humans , Laboratories/standards , Male , Postmenopause , Quality Control , Reproducibility of Results , Sensitivity and Specificity , Sex Characteristics
2.
Clin Lab ; 53(1-2): 1-9, 2007.
Article in English | MEDLINE | ID: mdl-17323819

ABSTRACT

BACKGROUND AND OBJECTIVE: Discrepancies between serum and heparin plasma samples have been described for many commercial troponin assays including the cardiac troponin T (cTnT) assay. Using the current 3rd generation Elecsys Troponin T immunoassay, heparin plasma cannot be recommended for the determination of cTnT due to systematic lower test results caused by a direct interference of the immunoassay by heparin. The purpose of the multicenter study was to evaluate the analytical performance of an improved 4th generation Elecsys Troponin T immunoassay with a special focus on the comparability of cTnT results determined in heparin plasma and serum. METHODS AND RESULTS: The multicenter evaluation was performed in 10 clinical laboratories according to a standardized protocol (Roche Diagnostics, Penzberg, Germany, Study No. B05P008). The Elecsys Troponin T immunoassay was performed on the Modular Analytics E170 and Elecsys 2010 systems. Intraassay imprecision (n = 21) and total imprecision (2 runs/d, 10 days, triplicate measurements) were evaluated using 2 commercial controls (Roche Diagnostics) and 6 different serum pools (cTnT: 0.0140 - 4.102 microg/L). Intraassay CVs ranged from 0.73 to 3.22%. Total imprecision CVs ranged from 3.61 to 35.45% (cTnT < 0.1 microg/L) and 1.82 to 9.09% (cTnT > 0.1 microg/L), respectively. The cut-off for myocardial necrosis was determined to be 0.03 microg/L using the 10% total imprecision CV criteria. Linearity was assessed by serial dilutions of 6 different serum samples using cTnT negative serum pools. Linearity was proven up to 21.3 microg/L (recoveries: 90% - 110%). Regression data of all comparison studies were calculated according to the method of Passing and Bablok. The method comparison between the 4th generation and the commercially available cTnT immunoassay showed highly similar results across the whole measuring range (0.01 - 25.0 microg/L): y = 1.024x -0.001, r = 0.998; n = 988. Using the commercially available cTnT reagent, the serum to heparin plasma comparison yielded a systematic bias to approximately 8% lower cTnT results in heparin plasma. However, suitable comparability was obtained using the 4th generation Elecsys cTnT assay. The regression analysis (serum vs. heparin plasma) across the studied measuring range (cTnT: 0.01 - 14 microg/L) yielded the following equation: y = 0.975x + 0.001; r = 0.986; n = 403. However, rare individual serum to matched heparin plasma samples still yielded poor comparability (deviation > 20%) using the 4th generation Elecsys Troponin T immunoassay. CONCLUSION: Our data confirm an excellent analytical performance of the improved troponin T immunoassay. Most importantly, no systematic bias between cTnT results determined in serum and heparin plasma was observed from data obtained in 7 evaluation sites. The performance of the 4th generation Elecsys Troponin T assay is therefore comparable to other commercially available troponin immunoassays. Further studies are necessary to investigate the cause of poor comparability of cTnT results in rare individual serum to matched heparin plasma samples.


Subject(s)
Immunoassay/instrumentation , Immunoassay/methods , Troponin T/analysis , Evaluation Studies as Topic , Heparin/analysis , Heparin/blood , Humans , Regression Analysis , Troponin T/blood
3.
J Trace Elem Med Biol ; 15(2-3): 73-8, 2001.
Article in English | MEDLINE | ID: mdl-11787990

ABSTRACT

An imbalance in the antioxidative system was connected with the development of a number of pathological processes. In order to receive values of a healthy group and to evaluate pathological changes of the trace element dependent antioxidative status in future, we investigated 99 healthy volunteers (45 male and 54 female, mean age 37.4 +/- 11.7 years). We determined the concentrations of Se, Cu and Zn, the concentrations of malondialdehyde (MDA) and the activities of the Se dependent glutathione peroxidase (GSH-Px) and the Zn/Cu dependent superoxide dismutase (SOD). The plasma concentrations (mean +/- SD) for Se, Cu and Zn were 0.84 +/- 0.10 micromol/l, 15.6 +/- 2.78 micromol/l and 12.6 +/- 1.80 micromol/l, resp., and for non protein-bound and protein bound MDA 0.27 +/- 0.07 micromol/l and 1.11 +/- 0.25 micromol/l, resp. The activity of GSH-Px in plasma and erythrocytes was 130 +/- 20.8 U/l and 19.8 +/- 4.18 U/mg Hb, resp. and of SOD in erythrocytes 3,159 +/- 847.2 U/g Hb. In plasma positive correlations have been found between Se concentrations and GSH-Px activities (p < 0.002, r = 0.31) and between GSH-Px activities and concentrations of non protein-bound MDA (p = 0.004, r = 0.28). A negative correlation has been observed between GSH-Px activities in plasma and in erythrocytes. The higher the concentrations of Cu in erythrocytes, the higher were the activities of SOD (p = 0.03, r = 0.22) and GSH-Px in erythrocytes (r = 0.26, p = 0.01), while an increasing activity of GSH-Px in these cells correlated with a decreasing concentration of non protein-bound MDA (r = -0,31, p = 0.002). An increase in BMI was connected with an increase in protein-bound MDA and a decrease in GSH-Px activities in pLasma (p = 0.002 and r = 0.23). As the results demonstrate, Se and Cu concentrations in erythrocytes can improve the trace element dependent antioxidative status.


Subject(s)
Antioxidants/pharmacology , Copper/chemistry , Selenium/chemistry , Trace Elements , Zinc/chemistry , Copper/metabolism , Erythrocytes/enzymology , Female , Glutathione Peroxidase/metabolism , Humans , Male , Malondialdehyde/chemistry , Superoxide Dismutase/metabolism