Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters










Database
Language
Publication year range
1.
J Cardiovasc Electrophysiol ; 26(8): 832-839, 2015 Aug.
Article in English | MEDLINE | ID: mdl-25917655

ABSTRACT

INTRODUCTION: There are limited comparative data on catheter ablation of atrial fibrillation (CAAF) using the second-generation cryoballoon (CB-2) versus point-by-point radiofrequency (RF). This study examines the acute/long-term CAAF outcomes using these 2 strategies. METHODS AND RESULTS: In this multicenter, retrospective, nonrandomized analysis, procedural and clinical outcomes of 1,196 patients (76% with paroxysmal AF) undergoing CAAF using CB-2 (n = 773) and open-irrigated, non-force sensing RF (n = 423) were evaluated. Pulmonary vein isolation was achieved in 98% with CB-2 and 99% with RF (P = 0.168). CB-2 was associated with shorter ablation time (40 ± 14 min vs. 66 ± 26 min; P < 0.001) and procedure time (145 ± 49 minutes vs. 188 ± 42 minutes; P < 0.001), but greater fluoroscopic utilization (29 ± 13 minutes vs. 23 ± 14 minutes; P < 0.001). While transient (7.6% vs. 0%; P < 0.001) and persistent (1.2% vs. 0%; P = 0.026) phrenic nerve palsy occurred exclusively with CB-2, other adverse event rates were similar between CB-2 (1.6%) and RF (2.6%); P = 0.207. However, freedom from AF/atrial flutter/tachycardia at 12 months following a single procedure without antiarrhythmic therapy was greater with CB-2 (76.6%) versus RF (60.4%); P < 0.001. While this difference was evident in patients with paroxysmal AF (P < 0.001), it did not reach significance in those with persistent AF (P = 0.089). Additionally, CB-2 was associated with reduced long-term need for antiarrhythmic therapy (16.7% vs. 22.0%; P = 0.024) and repeat ablations (14.6% vs. 24.1%; P < 0.001). CONCLUSION: In this multicenter, retrospective, nonrandomized study, CAAF using CB-2 coupled with RF as occasionally required was associated with greater freedom from atrial arrhythmias at 12 months following a single procedure without antiarrhythmic therapy when compared to open-irrigated, non-force sensing RF, alone.


Subject(s)
Atrial Fibrillation/surgery , Catheter Ablation , Cryosurgery/instrumentation , Pulmonary Veins/surgery , Therapeutic Irrigation , Action Potentials , Aged , Atrial Fibrillation/diagnosis , Atrial Fibrillation/physiopathology , Atrial Flutter/etiology , Brazil , Catheter Ablation/adverse effects , Cryosurgery/adverse effects , Disease-Free Survival , Equipment Design , Female , Heart Rate , Humans , Kaplan-Meier Estimate , Male , Middle Aged , Ontario , Operative Time , Pulmonary Veins/physiopathology , Radiography, Interventional , Recurrence , Retrospective Studies , Risk Factors , Tachycardia, Supraventricular/etiology , Therapeutic Irrigation/adverse effects , Time Factors , Treatment Outcome , United States
SELECTION OF CITATIONS
SEARCH DETAIL
...