1.
Appl Radiat Isot ; 205: 111160, 2024 Mar.
Article in English | MEDLINE | ID: mdl-38171076

ABSTRACT

Source strength measurements are of critical importance for brachytherapy and are often performed in hospitals using well-type chambers. Quality assurance and quality control procedures should be implemented, but that may prove difficult in some hospitals due to the lack of equipment or metrological support. A study was conducted to investigate the status of the measurement equipment in Serbian hospitals and to organize an intercomparison campaign using a hospital 192Ir source. All the hospitals were able to measure the source strength with the required accuracy, but the quality assurance can be improved. Two hospitals performed stability checks, which were evaluated. Uncertainty budget was created during the exercise and is presented in this paper. The described methodology can be used to quickly assess the performance of brachytherapy centers, and can be easily adapted to calibration procedure.


Subject(s)
Brachytherapy , Brachytherapy/methods , Serbia , Radiotherapy Dosage , Quality Control , Calibration , Radiometry/methods , Iridium Radioisotopes
SELECTION OF CITATIONS