Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters










Database
Language
Publication year range
1.
Opt Express ; 25(6): 7150-7160, 2017 Mar 20.
Article in English | MEDLINE | ID: mdl-28381054

ABSTRACT

Phase-shifting is one of the most useful methods of phase recovery in digital interferometry in the estimation of small displacements, but miscalibration errors of the phase shifters are very common. In practice, the main problem associated with such errors is related to the response of the phase shifter devices, since they are dependent on mechanical and/or electrical parts. In this work, a novel technique to detect and measure calibration errors in phase-shifting interferometry, when an unexpected phase shift arises, is proposed. The described method uses the Radon transform, first as an automatic-calibrating technique, and then as a profile measuring procedure when analyzing a specific zone of an interferogram. After, once maximum and minimum value parameters have been registered, these can be used to measure calibration errors. Synthetic and real interferograms are included in the testing, which has thrown good approximations for both cases, notwithstanding the interferogram fringe distribution or its phase-shifting steps. Tests have shown that this algorithm is able to measure the deviations of the steps in phase-shifting interferometry. The developed algorithm can also be used as an alternative in the calibration of phase shifter devices.

SELECTION OF CITATIONS
SEARCH DETAIL
...