Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 2 de 2
Filter
Add more filters










Database
Publication year range
1.
Rev Esp Anestesiol Reanim ; 57(2): 103-8, 2010 Feb.
Article in Spanish | MEDLINE | ID: mdl-20337002
2.
Rev Chir Orthop Reparatrice Appar Mot ; 85(3): 302-5, 1999 Jun.
Article in French | MEDLINE | ID: mdl-10422137

ABSTRACT

We report a case of dissociation between polyethylene and metal-backing of the acetabular component of a total hip prosthesis after reduction of a late dislocation. We refer to advantages and disadvantages of an elevated-rim acetabular liner in total hip arthroplasty. The position of the metallic markers of the polyethylene and/or excentric situation of the neck-head complex from the metal-back allowed us to diagnose it. An appropriate surgical technique avoiding the joint laxity, firm assembly of the modular components when they are implanted, gentle reduction of an eventual dislocation and fluoroscopic control, are factors which can avoid this rare complication.


Subject(s)
Hip Prosthesis/adverse effects , Joint Instability/diagnostic imaging , Joint Instability/etiology , Prosthesis Failure , Female , Fluoroscopy , Humans , Intraoperative Care/methods , Joint Instability/prevention & control , Manipulation, Orthopedic/methods , Middle Aged , Polyethylenes , Prosthesis Design