1.
Med Phys ; 37(9): 4817-53, 2010 Sep.
Article in English | MEDLINE | ID: mdl-20964201

ABSTRACT

Helical tomotherapy is a relatively new modality with integrated treatment planning and delivery hardware for radiation therapy treatments. In view of the uniqueness of the hardware design of the helical tomotherapy unit and its implications in routine quality assurance, the Therapy Physics Committee of the American Association of Physicists in Medicine commissioned Task Group 148 to review this modality and make recommendations for quality assurance related methodologies. The specific objectives of this Task Group are: (a) To discuss quality assurance techniques, frequencies, and tolerances and (b) discuss dosimetric verification techniques applicable to this unit. This report summarizes the findings of the Task Group and aims to provide the practicing clinical medical physicist with the insight into the technology that is necessary to establish an independent and comprehensive quality assurance program for a helical tomotherapy unit. The emphasis of the report is to describe the rationale for the proposed QA program and to provide example tests that can be performed, drawing from the collective experience of the task group members and the published literature. It is expected that as technology continues to evolve, so will the test procedures that may be used in the future to perform comprehensive quality assurance for helical tomotherapy units.


Subject(s)
Advisory Committees , Radiotherapy/standards , Research , Societies, Scientific , Calibration , Health Planning Guidelines , Humans , Quality Control , Radiometry , Radiotherapy Planning, Computer-Assisted , Tomography, X-Ray Computed
2.
Med Phys ; 36(11): 5359-73, 2009 Nov.
Article in English | MEDLINE | ID: mdl-19994544

ABSTRACT

AAPM Task Group 119 has produced quantitative confidence limits as baseline expectation values for IMRT commissioning. A set of test cases was developed to assess the overall accuracy of planning and delivery of IMRT treatments. Each test uses contours of targets and avoidance structures drawn within rectangular phantoms. These tests were planned, delivered, measured, and analyzed by nine facilities using a variety of IMRT planning and delivery systems. Each facility had passed the Radiological Physics Center credentialing tests for IMRT. The agreement between the planned and measured doses was determined using ion chamber dosimetry in high and low dose regions, film dosimetry on coronal planes in the phantom with all fields delivered, and planar dosimetry for each field measured perpendicular to the central axis. The planar dose distributions were assessed using gamma criteria of 3%/3 mm. The mean values and standard deviations were used to develop confidence limits for the test results using the concept confidence limit = /mean/ + 1.96sigma. Other facilities can use the test protocol and results as a basis for comparison to this group. Locally derived confidence limits that substantially exceed these baseline values may indicate the need for improved IMRT commissioning.


Subject(s)
Radiometry , Radiotherapy Planning, Computer-Assisted , Radiotherapy, Intensity-Modulated/standards , Film Dosimetry , Head and Neck Neoplasms/radiotherapy , Humans , Male , Phantoms, Imaging , Prostatic Neoplasms/radiotherapy , Quality Assurance, Health Care
3.
Med Phys ; 34(8): 3249-62, 2007 Aug.
Article in English | MEDLINE | ID: mdl-17879788

ABSTRACT

Quality assurance (QA) is a topic of major concern in the field of intensity modulated radiation therapy (IMRT). The standard of practice for IMRT is to perform QA testing for individual patients to verify that the dose distribution will be delivered to the patient. The purpose of this study was to develop a new technique that could eventually be used to automatically evaluate helical tomotherapy treatments during delivery using exit detector data. This technique uses an autoassociative kernel regression (AAKR) model to detect errors in tomotherapy delivery. AAKR is a novel nonparametric model that is known to predict a group of correct sensor values when supplied a group of sensor values that is usually corrupted or contains faults such as machine failure. This modeling scheme is especially suited for the problem of monitoring the fluence values found in the exit detector data because it is able to learn the complex detector data relationships. This scheme still applies when detector data are summed over many frames with a low temporal resolution and a variable beam attenuation resulting from patient movement. Delivery sequences from three archived patients (prostate, lung, and head and neck) were used in this study. Each delivery sequence was modified by reducing the opening time for random individual multileaf collimator (MLC) leaves by random amounts. The errof and error-free treatments were delivered with different phantoms in the path of the beam. Multiple autoassociative kernel regression (AAKR) models were developed and tested by the investigators using combinations of the stored exit detector data sets from each delivery. The models proved robust and were able to predict the correct or error-free values for a projection, which had a single MLC leaf decrease its opening time by less than 10 msec. The model also was able to determine machine output errors. The average uncertainty value for the unfaulted projections ranged from 0.4% to 1.8% of the detector signal. The low model uncertainty indicates that the AAKR model is extremely accurate in its predictions and also suggests that the model may be able to detect errors that cause the fluence to change by less than 2%. However, additional evaluation of the AAKR technique is needed to determine the minimum detectable error threshold from the compressed helical tomotherapy detector data. Further research also needs to explore applying this technique to electronic portal imaging detector data.


Subject(s)
Neoplasms/pathology , Radiotherapy Planning, Computer-Assisted/methods , Radiotherapy, Intensity-Modulated/instrumentation , Radiotherapy, Intensity-Modulated/methods , Tomography, X-Ray Computed/methods , Automation , Calibration , Humans , Image Processing, Computer-Assisted , Models, Statistical , Neoplasms/diagnostic imaging , Normal Distribution , Quality Control , Regression Analysis , Reproducibility of Results , Tomography Scanners, X-Ray Computed , Xenon
4.
Med Phys ; 34(8): 3286-93, 2007 Aug.
Article in English | MEDLINE | ID: mdl-17879792

ABSTRACT

Helical tomotherapy is a treatment technique that is delivered from a 6 MV fan beam that traces a helical path while the couch moves linearly into the bore. In order to increase the treatment delivery dose rate, helical tomotherapy systems do not have a flattening filter. As such, the dose distributions near the surface of the patient may be considerably different from other forms of intensity-modulated delivery. The purpose of this study was to measure the dose distributions near the surface for helical tomotherapy plans with a varying separation between the target volume and the surface of an anthropomorphic phantom. A hypothetical planning target volume (PTV) was defined on an anthropomorphic head phantom to simulate a 2.0 Gy per fraction IMRT parotid-sparing head and neck treatment of the upper neck nodes. A total of six target volumes were created with 0, 1, 2, 3, 4, and 5 mm of separation between the surface of the phantom and the outer edge of the PTV. Superficial doses were measured for each of the treatment deliveries using film placed in the head phantom and thermoluminescent dosimeters (TLDs) placed on the phantom's surface underneath an immobilization mask. In the 0 mm test case where the PTV extends to the phantom surface, the mean TLD dose was 1.73 +/- 0.10 Gy (or 86.6 +/- 5.1% of the prescribed dose). The measured superficial dose decreases to 1.23 +/- 0.10 Gy (61.5 +/- 5.1% of the prescribed dose) for a PTV-surface separation of 5 mm. The doses measured by the TLDs indicated that the tomotherapy treatment planning system overestimates superficial doses by 8.9 +/- 3.2%. The radiographic film dose for the 0 mm test case was 1.73 +/- 0.07 Gy, as compared to the calculated dose of 1.78 +/- 0.05 Gy. Given the results of the TLD and film measurements, the superficial calculated doses are overestimated between 3% and 13%. Without the use of bolus, tumor volumes that extend to the surface may be underdosed. As such, it is recommended that bolus be added for these clinical cases. For cases where the target volume is located 1 to 5 mm below the surface, the tumor volume coverage can be achieved with surface doses ranging from 56% to 93% of the prescribed dose.


Subject(s)
Head/diagnostic imaging , Head/pathology , Tomography, X-Ray Computed/methods , Dose-Response Relationship, Radiation , Humans , Models, Statistical , Phantoms, Imaging , Radiometry/methods , Radiotherapy Planning, Computer-Assisted/methods , Radiotherapy, Intensity-Modulated/methods , Thermoluminescent Dosimetry , Time Factors , Tomography, X-Ray Computed/instrumentation
5.
J Appl Clin Med Phys ; 8(3): 37-51, 2007 Jul 17.
Article in English | MEDLINE | ID: mdl-17712296

ABSTRACT

The purpose of the present study was to implement a technique for daily computed tomography (CT)-based image-guided radiation therapy and to report observations on treatment planning, imaging, and delivery based on the first 2 years of clinical experience. Patients with previously untreated stage T1-T3 biopsy-proven adenocarcinoma of the prostate were considered eligible for treatment with daily CT-guided helical tomotherapy. The prostate was targeted daily using megavoltage CT (MVCT) images that were fused with treatment-planning CT images based on anatomic alignments. All patients were treated at 2 Gy per fraction to 76-78 Gy (mean: 76.7 Gy). As part of this study, 33 prostate patients were planned, imaged, and treated with a total of 1266 CT-guided fractions. The prostate, rectum, bladder, femoral heads, and pubis symphysis were visible in one or more slices for all 1266 MVCT image sets. The typical range of measured prostate displacement relative to a 3-point external laser setup in this study was 2-10 mm [3.4 mm standard deviation (SD)] in the anterior-posterior direction, 2-8 mm (3.7 mm SD) in the lateral direction, and 1-6 mm (2.4 mm SD) in the superior-inferior direction. The obese patients in this study had a substantially larger lateral variation (8.2 mm SD) attributable to mobility of skin marks. The prostate, seminal vesicles, rectum, and bladder anatomy were used to position the patient relative to the desired treatment position without the use of implanted markers. Acute toxicities were within the expected range given the number of patients treated and the dose level.


Subject(s)
Adenocarcinoma/diagnostic imaging , Adenocarcinoma/radiotherapy , Prostatic Neoplasms/diagnostic imaging , Prostatic Neoplasms/radiotherapy , Radiotherapy, Computer-Assisted/methods , Radiotherapy, Conformal/methods , Tomography, Spiral Computed/methods , Humans , Male , Pilot Projects , Treatment Outcome
6.
Int J Radiat Oncol Biol Phys ; 67(2): 601-9, 2007 Feb 01.
Article in English | MEDLINE | ID: mdl-17236977

ABSTRACT

PURPOSE: Volumetric computed tomography (CT) images acquired by image-guided radiation therapy (IGRT) systems can be used to measure tumor response over the course of treatment. Predictive adaptive therapy is a novel treatment technique that uses volumetric IGRT data to actively predict the future tumor response to therapy during the first few weeks of IGRT treatment. The goal of this study was to develop and test a model for predicting lung tumor response during IGRT treatment using serial megavoltage CT (MVCT). METHODS AND MATERIALS: Tumor responses were measured for 20 lung cancer lesions in 17 patients that were imaged and treated with helical tomotherapy with doses ranging from 2.0 to 2.5 Gy per fraction. Five patients were treated with concurrent chemotherapy, and 1 patient was treated with neoadjuvant chemotherapy. Tumor response to treatment was retrospectively measured by contouring 480 serial MVCT images acquired before treatment. A nonparametric, memory-based locally weight regression (LWR) model was developed for predicting tumor response using the retrospective tumor response data. This model predicts future tumor volumes and the associated confidence intervals based on limited observations during the first 2 weeks of treatment. The predictive accuracy of the model was tested using a leave-one-out cross-validation technique with the measured tumor responses. RESULTS: The predictive algorithm was used to compare predicted verse-measured tumor volume response for all 20 lesions. The average error for the predictions of the final tumor volume was 12%, with the true volumes always bounded by the 95% confidence interval. The greatest model uncertainty occurred near the middle of the course of treatment, in which the tumor response relationships were more complex, the model has less information, and the predictors were more varied. The optimal days for measuring the tumor response on the MVCT images were on elapsed Days 1, 2, 5, 9, 11, 12, 17, and 18 during treatment. CONCLUSIONS: The LWR model accurately predicted final tumor volume for all 20 lung cancer lesions. These predictions were made using only 8 days' worth of observations from early in the treatment. Because the predictions are accurate with quantified uncertainty, they could eventually be used to optimize treatment.


Subject(s)
Algorithms , Lung Neoplasms/radiotherapy , Models, Biological , Tomography, Spiral Computed , Aged , Aged, 80 and over , Confidence Intervals , Dose Fractionation, Radiation , Humans , Lung Neoplasms/diagnostic imaging , Lung Neoplasms/drug therapy , Middle Aged , Predictive Value of Tests , Radiotherapy Planning, Computer-Assisted , Retrospective Studies , Treatment Outcome , Uncertainty
7.
Med Phys ; 33(10): 3874-900, 2006 Oct.
Article in English | MEDLINE | ID: mdl-17089851

ABSTRACT

This document is the report of a task group of the AAPM and has been prepared primarily to advise medical physicists involved in the external-beam radiation therapy of patients with thoracic, abdominal, and pelvic tumors affected by respiratory motion. This report describes the magnitude of respiratory motion, discusses radiotherapy specific problems caused by respiratory motion, explains techniques that explicitly manage respiratory motion during radiotherapy and gives recommendations in the application of these techniques for patient care, including quality assurance (QA) guidelines for these devices and their use with conformal and intensity modulated radiotherapy. The technologies covered by this report are motion-encompassing methods, respiratory gated techniques, breath-hold techniques, forced shallow-breathing methods, and respiration-synchronized techniques. The main outcome of this report is a clinical process guide for managing respiratory motion. Included in this guide is the recommendation that tumor motion should be measured (when possible) for each patient for whom respiratory motion is a concern. If target motion is greater than 5 mm, a method of respiratory motion management is available, and if the patient can tolerate the procedure, respiratory motion management technology is appropriate. Respiratory motion management is also appropriate when the procedure will increase normal tissue sparing. Respiratory motion management involves further resources, education and the development of and adherence to QA procedures.


Subject(s)
Neoplasms/radiotherapy , Radiation Oncology/methods , Radiotherapy Planning, Computer-Assisted/methods , Radiotherapy, Intensity-Modulated/methods , Respiration , Guidelines as Topic , Humans , Motion , Neoplasms/pathology , Quality Control , Radiometry , Radiotherapy Dosage , Radiotherapy, Computer-Assisted , Time Factors , Tomography, X-Ray Computed/methods
8.
Med Phys ; 33(8): 2751-6, 2006 Aug.
Article in English | MEDLINE | ID: mdl-16964850

ABSTRACT

Topographic treatment is a radiation therapy delivery technique for fixed-gantry (nonrotational) treatments on a helical tomotherapy system. The intensity-modulated fields are created by moving the treatment couch relative to a fan-beam positioned at fixed gantry angles. The delivered dose distribution is controlled by moving multileaf collimator (MLC) leaves into and out of the fan beam. The purpose of this work was to develop a leaf-sequencing algorithm for creating topographic MLC sequences. Topographic delivery was modeled using the analogy of a water faucet moving over a collection of bottles. The flow rate per unit length of the water from the faucet represented the photon fluence per unit length along the width of the fan beam, the collection of bottles represented the pixels in the treatment planning fluence map, and the volume of water collected in each bottle represented the delivered fluence. The radiation fluence per unit length delivered to the target at a given position is given by the convolution of the intensity distribution per unit length over the width of the beam and the time per unit distance along the direction of travel that an MLC leaf is open. The MLC opening times for the desired dose profiles were determined using a technique based on deconvolution using a genetic algorithm. The MLC opening times were expanded in terms of a Fourier series, and a genetic algorithm was used to find the best expansion coefficients for a given dose distribution. A series of wedge shapes (15, 30, 45, and 60 deg) and "dose well" test fluence maps were created to test the algorithm's ability to generate topographic leaf sequences. The accuracy of the leaf-sequencing algorithm was measured on a helical tomotherapy system using radiographic film placed at depth in water equivalent material. The measured dose profiles were compared with the desired dose distributions. The agreement was within +/- 2% or 2 mm distance-to-agreement (DTA) in the high dose gradient regions for all test cases. The central axis measured dose was between 3.6% and 4.2% higher than the expected dose for the wedge cases. For the "dose well" test cases, the calculated and measured doses agreed to within +/- 0.5% at the peak and within +/- 1.6% in the "dose well." The topographic leaf-sequencing algorithm produced deliverable dose distributions that agreed well with the calculated dose distributions. This delivery technique could be used for treatment of whole intact breast. However, additional work is needed to further improve the algorithm in order to get better agreement between the calculated, deliverable, and measured dose distributions.


Subject(s)
Algorithms , Models, Biological , Radiometry/methods , Radiotherapy Planning, Computer-Assisted/methods , Radiotherapy, Computer-Assisted/methods , Radiotherapy, Conformal/instrumentation , Radiotherapy, Conformal/methods , Computer Simulation , Humans , Radiotherapy Dosage , Radiotherapy Planning, Computer-Assisted/instrumentation , Radiotherapy, Computer-Assisted/instrumentation , Reproducibility of Results , Sensitivity and Specificity
9.
Int J Radiat Oncol Biol Phys ; 64(4): 1237-44, 2006 Mar 15.
Article in English | MEDLINE | ID: mdl-16446055

ABSTRACT

PURPOSE: The gross tumor volume (GTV) for many lung cancer patients can decrease during the course of radiation therapy. As the tumor reduces in size during treatment, the margin added around the GTV effectively becomes larger, which can result in the excessive irradiation of normal lung tissue. The specific goal of this study is to evaluate the feasibility of using image-guided adaptive radiation therapy to adjust the planning target volume weekly based on the previous week's CT image sets that were used for image-guided patient setup. METHODS AND MATERIALS: Megavoltage computed tomography (MVCT) images of the GTV were acquired daily on a helical tomotherapy system. These images were used to position the patient and to measure reduction in GTV volume. A planning study was conducted to determine the amount of lung-sparing that could have been achieved if adaptive therapy had been used. Treatment plans were created in which the target volumes were reduced after tumor reduction was measured. RESULTS: A total of 158 MVCT imaging sessions were performed on 7 lung patients. The GTV was reduced by 60-80% during the course of treatment. The tumor reduction in the first 60 days of treatment can be modeled using the second-order polynomial R = 0.0002t(2) - 0.0219t + 1.0, where R is the percent reduction in GTV, and t is the number of elapsed days. Based on these treatment planning studies, the absolute volume of ipsilateral lung receiving 20 Gy can be reduced between 17% and 23% (21% mean) by adapting the treatment delivery. The benefits of adaptive therapy are the greatest for tumor volumes > or =25 cm3 and are directly dependent on GTV reduction during treatment. CONCLUSIONS: Megavoltage CT-based image guidance can be used to position lung cancer patients daily. This has the potential to decrease margins associated with daily setup error. Furthermore, the adaptive therapy technique described in this article can decrease the volume of healthy lung tissue receiving above 20 Gy. However, further study is needed to determine whether adaptive therapy could result in the underdosing of microscopic extension.


Subject(s)
Carcinoma, Non-Small-Cell Lung/diagnostic imaging , Carcinoma, Non-Small-Cell Lung/radiotherapy , Lung Neoplasms/diagnostic imaging , Lung Neoplasms/radiotherapy , Tomography, Spiral Computed/methods , Carcinoma, Non-Small-Cell Lung/pathology , Feasibility Studies , Humans , Lung Neoplasms/pathology , Radiology Information Systems , Remission Induction , Time Factors
10.
Int J Radiat Oncol Biol Phys ; 63(5): 1576-83, 2005 Dec 01.
Article in English | MEDLINE | ID: mdl-16125871

ABSTRACT

INTRODUCTION: Patients with vertebral metastasis that receive radiation therapy are typically treated to the spinal cord tolerance dose. As such, it is difficult to successfully deliver a second course of radiation therapy for patients with overlapping treatment volumes. In this study, an image-guided helical tomotherapy system was evaluated for the retreatment of previously irradiated vertebral metastasis. METHODS AND MATERIALS: Helical tomotherapy dose gradients and maximum cord doses were measured in a cylindrical phantom for geometric test cases with separations between the planning target volume (PTV) and the spinal cord organ at risk (OAR) of 2 mm, 4 mm, 6 mm, 8 mm, and 10 mm. Megavoltage computed tomography (CT) images were examined for their ability to localize spinal anatomy for positioning purposes by repeat imaging of the cervical spine in an anthropomorphic phantom. In addition to the phantom studies, 8 patients with cord compressions that had received previous radiation therapy were retreated to a mean dose of 28 Gy using conventional fractionation. RESULTS AND DISCUSSION: Megavoltage CT images were capable of positioning an anthropomorphic phantom to within +/-1.2 mm (2sigma) superior-inferiorly and within +/-0.6 mm (2sigma) anterior-posteriorly and laterally. Dose gradients of 10% per mm were measured in phantom while PTV uniformity indices of less than 11% were maintained. The calculated maximum cord dose was 25% of the prescribed dose for a 10-mm PTV-to-OAR separation and 71% of the prescribed dose for a PTV-to-OAR separation of 2 mm. Eight patients total have been treated without radiation-induced myelopathy or any other adverse effects from treatment. CONCLUSIONS: A technique has been evaluated for the retreatment of vertebral metastasis using image-guided helical tomotherapy. Phantom and patient studies indicated that a tomotherapy system is capable of delivering dose gradients of 10% per mm and positioning the patient within 1.2 mm without the use of special stereotactic immobilization.


Subject(s)
Phantoms, Imaging , Spinal Neoplasms/radiotherapy , Spinal Neoplasms/secondary , Tomography, Spiral Computed/methods , Adult , Aged , Humans , Middle Aged , Radiotherapy Planning, Computer-Assisted/methods
11.
Med Phys ; 31(7): 2119-20, 2004 Jul.
Article in English | MEDLINE | ID: mdl-15305465

ABSTRACT

The output and energy calibrations for the first clinical Hi-Art 2.0 helical tomotherapy system have been reviewed. Fixed-gantry/fixed-couch and rotational-gantry/fixed-couch measurements were made on a daily basis over a period of 20 weeks to investigate system stability. Static gantry measurements were taken at 10 cm depth in a rectangular stack of Virtual Water at an SSD distance of 90 cm and a field size of 5 x 40 cm. Rotational gantry measurements were taken in a cylindrical phantom Virtual Water phantom for a field size of 5 x 40 cm. The Hi-Art 2.0 system has maintained its calibration to within +/-2% and energy to within +/- 1.5% over the initial 20 week period.


Subject(s)
Equipment Failure Analysis/methods , Quality Assurance, Health Care/methods , Radiometry/methods , Radiotherapy, Conformal/instrumentation , Calibration/standards , Equipment Failure Analysis/standards , Linear Energy Transfer , Quality Assurance, Health Care/standards , Radiometry/standards , Radiotherapy Dosage/standards , Radiotherapy, Conformal/standards , Reference Standards , Reproducibility of Results , Sensitivity and Specificity