1.
Future Cardiol ; 17(7): 1193-1197, 2021 10.
Article in English | MEDLINE | ID: mdl-33448229

ABSTRACT

Background: Vascular closure devices have replaced mechanical compression for closure of femoral access sites after endovascular procedures. Case: We present an 87-year-old male with symptomatic infrarenal aortic aneurysm measuring 4.8 cm presenting for elective endovascular repair of the aortic aneurysm. A Perclose ProGlide Suture-Mediated Closure was used for closure. The closure was complicated by a separation of the ProGlide device resulting in the migration of the footplate to the descending aorta. Correction required snare retrieval via radial access, and the patient recovered without complications. Discussion: We highlight an important complication of the Perclose ProGlide Suture-Mediated Closure device that is rare but important to know when performing endovascular closures with this device.


Subject(s)
Endovascular Procedures , Vascular Closure Devices , Aged, 80 and over , Femoral Artery/surgery , Humans , Male , Punctures , Treatment Outcome