Int J Lab Hematol ; 39(1): 95-100, 2017 Feb.
Article in English | MEDLINE | ID: mdl-27997764

ABSTRACT

INTRODUCTION: Hemoglobin (Hb) concentration is often measured in global health and nutrition surveys to determine anemia prevalence using a portable hemoglobinometer such as the Hemocue® Hb 201+. More recently, a newer model was released (Hemocue Hb 301) utilizing slightly different methods to measure Hb as compared to the older model. The objective was to measure bias and concordance between Hb concentrations using the Hemocue Hb 301 and Hb 201+ models in a rural field setting. METHODS: Hemoglobin (Hb) concentration was measured using one finger prick of blood (approximately 10 µL) from 175 Cambodian women (18-49 years) using three Hemocue Hb 201+ and three Hb 301 machines. Bias and concordance were measured and plotted. RESULTS: Overall, mean ± SD Hb concentration was 116 ± 13 g/L using the Hb 201+ and 118 ± 12 g/L using the Hb 301; and anemia prevalence (Hb < 120 g/L) was 58% (n = 102) and 58% (n = 101), respectively. Overall bias ± SD was 2.0 ± 10.5 g/L and concordance (95% CI) was 0.63 (0.54, 0.72). Despite the 2 g/L bias detected between models, anemia prevalence was very similar in both models. CONCLUSIONS: The two models measured anemia prevalence comparably in this population of women in rural Cambodia.


Subject(s)
Anemia/blood , Hemoglobins/metabolism , Anemia/epidemiology , Cambodia/epidemiology , Female , Hemoglobinometry/instrumentation , Humans , Prevalence