ABSTRACT
Beta Secondary Standard 2 (BSS 2) provides beta radiation fields with certified values of absorbed dose to tissue and the derived operational radiation protection quantities. As part of the quality assurance, the reliability of the CDTN BSS2 system was verified through measurements in the 90Sr/90Y and 85Kr beta radiation fields. Absorbed dose rates and their angular variation were measured with a 23392 model PTW extrapolation chamber and with Gafchromic radiochromic films on a PMMA slab phantom. The feasibility of using both methods was analyzed.
ABSTRACT
The feasibility of using radiochromic films to verify the metrological coherence among standard beta radiation fields was evaluated. Exercises were done between two Brazilian metrology laboratories in beta fields from (90)Sr/(90)Y, (85)Kr and (147)Pm radiation sources. Results showed that the radiochromic film was useful for field mapping aiming uniformity and alignment verification and it was not reliable for absorbed dose measurements only for (147)Pm beta field.