Circ Arrhythm Electrophysiol ; : e012635, 2024 May 28.
Article in English | MEDLINE | ID: mdl-38804141

ABSTRACT

BACKGROUND: Irrigated radiofrequency ablation with half-normal saline can potentially increase lesion size but may increase the risk of steam pops with the risk of emboli or perforation. We hypothesized that pops would be preceded by intracardiac echocardiography (ICE) findings as well as a large impedance fall. METHODS: In 100 consecutive patients undergoing endocardial ventricular arrhythmia radiofrequency ablation with half-normal saline, we attempted to observe the ablation site with ICE. Radiofrequency ablation power was titrated to a 15 to 20 Ohm impedance fall and could be adjusted for tissue whitening and increasing bubble formation on ICE. Steam pops were defined as audible or a sudden explosion of microbubbles on ICE. RESULTS: Of 2190 ablation applications in 100 patients (82% cardiomyopathy, 50% sustained ventricular tachycardia), pops occurred during 43 (2.0%) applications. Sites with pops had greater impedance decreases of 18 [14, 21]% versus 13 [10, 17]% (P<0.001). ICE visualized 1308 (59.7%) radiofrequency sites, and fewer pops occurred when ICE visualized the radiofrequency ablation site (1.4%) compared with without ICE visualization (2.8%; P=0.016). Of the 18 ICE-visible pops, 7 (39%) were silent but recognized as an explosion of bubbles on ICE. With ICE, 89% of pops were preceded by either tissue whitening or a sudden increase in bubbles. In a multivariable model, tissue whitening and a sudden increase in bubbles were associated with steam pops (odds ratio, 7.186; P=0.004, and odds ratio, 29.93; P<0.001, respectively), independent of impedance fall and power. There were no pericardial effusions or embolic events with steam pops. CONCLUSIONS: Steam pops occurred in 2% of half-normal saline radiofrequency applications titrated to an impedance fall and are likely under-recognized without ICE. On ICE, steam pops are usually preceded by tissue whitening or a sudden increase in bubble formation, which can potentially be used to adjust radiofrequency application to help reduce pops.

3.
J Cardiovasc Electrophysiol ; 35(5): 994-1004, 2024 May.
Article in English | MEDLINE | ID: mdl-38501333

ABSTRACT

INTRODUCTION: When ventricular tachycardia (VT) recurs after standard RF ablation (sRFA) some patients benefit from repeat sRFA, whereas others warrant advanced methods such as intramural needle ablation (INA). Our objectives are to assess the utility of repeat sRFA and to clarify the benefit of INA when repeat sRFA fails in patients with VT due to structural heart disease. METHODS: In consecutive patients who were prospectively enrolled in a study for INA for recurrent sustained monomorphic VT despite sRFA, repeat sRFA was considered first. INA was performed during the same procedure if repeat sRFA failed or no targets for sRFA were identified. RESULTS: Of 85 patients enrolled, acute success with repeat sRFA was achieved in 30 patients (35%), and during the 6-month follow-up, 87% (20/23) were free of VT hospitalization, 78% were free of any VT, and 7 were lost to follow-up. INA was performed in 55 patients (65%) after sRFA failed, or no endocardial targets were found abolished or modified inducible VT in 35/55 patients (64%). During follow-up, 72% (39/54) were free of VT hospitalization, 41% were free of any VT, and 1 was lost to follow-up. Overall, 59 out of 77 (77%) patients were free of hospitalization and 52% were free of any VT. Septal-origin VTs were more likely to need INA, whereas RV and papillary muscle VTs were less likely to require INA. CONCLUSIONS: Repeat sRFA was beneficial in 23% (18/77) of patients with recurrent sustained VT who were referred for INA. The availability of INA increased favorable outcomes to 52%.


Subject(s)
Catheter Ablation , Cicatrix , Recurrence , Reoperation , Tachycardia, Ventricular , Humans , Tachycardia, Ventricular/surgery , Tachycardia, Ventricular/physiopathology , Tachycardia, Ventricular/diagnosis , Tachycardia, Ventricular/etiology , Male , Female , Middle Aged , Aged , Prospective Studies , Catheter Ablation/adverse effects , Cicatrix/physiopathology , Cicatrix/diagnosis , Cicatrix/surgery , Cicatrix/etiology , Time Factors , Action Potentials , Needles , Heart Rate , Risk Factors , Treatment Outcome
4.
Heart Rhythm ; 21(6): 956-957, 2024 Jun.
Article in English | MEDLINE | ID: mdl-38431023
5.
J Am Heart Assoc ; 13(6): e031029, 2024 Mar 19.
Article in English | MEDLINE | ID: mdl-38471835

ABSTRACT

BACKGROUND: Recurrence after atrial fibrillation (AF) ablation remains common. We evaluated the association between recurrence and levels of biomarkers of cardiac remodeling, and their ability to improve recurrence prediction when added to a clinical prediction model. METHODS AND RESULTS: Blood samples collected before de novo catheter ablation were analyzed. Levels of bone morphogenetic protein-10, angiopoietin-2, fibroblast growth factor-23, insulin-like growth factor-binding protein-7, myosin-binding protein C3, growth differentiation factor-15, interleukin-6, N-terminal pro-brain natriuretic peptide, and high-sensitivity troponin T were measured. Recurrence was defined as ≥30 seconds of an atrial arrhythmia 3 to 12 months postablation. Multivariable logistic regression was performed using biomarker levels along with clinical covariates: APPLE score (Age >65 years, Persistent AF, imPaired eGFR [<60 ml/min/1.73m2], LA diameter ≥43 mm, EF <50%; which includes age, left atrial diameter, left ventricular ejection fraction, persistent atrial fibrillation, and estimated glomerular filtration rate), preablation rhythm, sex, height, body mass index, presence of an implanted continuous monitor, year of ablation, and additional linear ablation. A total of 1873 participants were included. A multivariable logistic regression showed an association between recurrence and levels of angiopoietin-2 (odds ratio, 1.08 [95% CI, 1.02-1.15], P=0.007) and interleukin-6 (odds ratio, 1.02 [95% CI, 1.003-1.03]; P=0.02). The area under the receiver operating characteristic curve of a model that only contained clinical predictors was 0.711. The addition of any of the 9 studied biomarkers to the predictive model did not result in a statistically significant improvement in the area under the receiver operating characteristic curve. CONCLUSIONS: Higher angiopoietin-2 and interleukin-6 levels were associated with recurrence after atrial fibrillation ablation in multivariable modeling. However, the addition of biomarkers to a clinical prediction model did not significantly improve recurrence prediction.


Subject(s)
Atrial Fibrillation , Atrial Remodeling , Catheter Ablation , Humans , Aged , Atrial Fibrillation/diagnosis , Atrial Fibrillation/surgery , Angiopoietin-2 , Interleukin-6 , Models, Statistical , Stroke Volume , Ventricular Remodeling , Risk Factors , Prognosis , Recurrence , Ventricular Function, Left , Biomarkers , Catheter Ablation/adverse effects , Catheter Ablation/methods , Treatment Outcome
6.
Europace ; 26(2)2024 Feb 01.
Article in English | MEDLINE | ID: mdl-38367008

ABSTRACT

AIMS: Failure of radiofrequency (RF) ablation of ventricular arrhythmias is often due to inadequate lesion size. Irrigated RF ablation with half-normal saline (HNS) has the potential to increase lesion size and reduce sodium delivery to the patient if the same volume of RF irrigant were used for normal saline (NS) and HNS but could increase risks related to steam pops and lesion size. This study aims to assess periprocedural complications and acute ablation outcome of ventricular arrhythmias ablation with HNS. METHODS AND RESULTS: Prospective assessment of outcomes was performed in 1024 endocardial and/or epicardial RF ablation procedures in 935 consecutive patients (median age 64 years, 71.2% men, 73.4% cardiomyopathy, 47.2% sustained ventricular tachycardia). Half-normal saline was selected at the discretion of the treating physician. Radiofrequency ablation power was generally titrated to a ≤15â€…Ω impedance fall with intracardiac echocardiography monitoring. Half-normal saline was used in 900 (87.9%) and NS in 124 (12.1%) procedures. Any adverse event within 30 days occurred in 13.0% of patients treated with HNS RF ablation including 4 (0.4%) strokes/transient ischaemic attacks and 34 (3.8%) pericardial effusions requiring treatment (mostly related to epicardial access). Two steam pops with perforation required surgical repair (0.2%). Patients who received NS irrigation had less severe disease and arrhythmias. In multivariable models, adverse events and acute success of the procedure were not related to the type of irrigation. CONCLUSION: Half-normal saline irrigation RF ablation with power guided by impedance fall and intracardiac echocardiography has an acceptable rate of complications and acute ablation success while administering half of the saline load expected for NS irrigation.


Subject(s)
Catheter Ablation , Radiofrequency Ablation , Tachycardia, Ventricular , Male , Humans , Middle Aged , Female , Saline Solution/adverse effects , Steam , Prospective Studies , Arrhythmias, Cardiac/diagnosis , Arrhythmias, Cardiac/surgery , Catheter Ablation/adverse effects , Catheter Ablation/methods , Tachycardia, Ventricular/surgery , Therapeutic Irrigation/adverse effects
7.
JACC Clin Electrophysiol ; 10(2): 193-202, 2024 Feb.
Article in English | MEDLINE | ID: mdl-38069975

ABSTRACT

BACKGROUND: Risks of radiofrequency catheter ablation for ventricular arrhythmias include emboli and bleeding complications but data on antithrombotic regimens are limited and guidelines do not specify a systematic approach. OBJECTIVES: This study sought to assess embolic and bleeding complications in relation to pre-periprocedure and post-periprocedure antithrombotic regimens. METHODS: Prospective assessment for complications was performed for 663 endocardial radiofrequency catheter ablation procedures in 616 consecutive patients (median age 64 years [Q1-Q3: 54-73 years], 70.3% men, 71.6% with cardiomyopathy, 44.5% with sustained ventricular tachycardia). RESULTS: There were 2 strokes (0.3%; 95% CI: 0.0%-0.8%), 1 transient ischemic attack (0.15%), and 2 pulmonary emboli (0.3%). There were 39 bleeding complications (5.9%) including 11 pericardial effusions (1.7%), and 28 related to vascular access (4.2%). Consistent with the prevalence of coronary artery disease (47.5%), atrial fibrillation (30.0%), and prior stroke (10.6%), preprocedure, 464 patients (70.0%) were taking antithrombotic agents including 220 (33.2%) taking aspirin alone (ASA), and 163 (24.6%) taking warfarin or a direct acting oral anticoagulant (DOAC). Preprocedure non-ASA antiplatelet use (OR: 2.846; P = 0.011) and DOAC use (OR: 2.585; P = 0.032) were associated with risk of bleeding complications. Following ablation, 49.8% of patients were treated with ASA 325 mg/d and 30.3% received DOACs or warfarin. New DOAC or warfarin administration was initiated in only 6.6% of patients. Overall, 39.7% of patients continued the same preprocedure antithrombotic regimen. CONCLUSIONS: Stroke is a rare complication of radiofrequency catheter ablation for ventricular arrhythmia using ASA 325 mg/d as a minimal postprocedure regimen with more potent regimens for selected patients.


Subject(s)
Atrial Fibrillation , Stroke , Male , Humans , Middle Aged , Female , Warfarin/adverse effects , Anticoagulants/adverse effects , Hemorrhage/etiology , Hemorrhage/chemically induced , Fibrinolytic Agents , Prospective Studies , Stroke/etiology , Stroke/epidemiology , Atrial Fibrillation/surgery , Aspirin/adverse effects
9.
J Cardiovasc Electrophysiol ; 34(11): 2403-2405, 2023 Nov.
Article in English | MEDLINE | ID: mdl-37787011

ABSTRACT

INTRODUCTION: Intracardiac echocardiography (ICE) reveals mobile thrombus on implantable electronic device leads in some patients undergoing electrophysiologic procedures. METHODS: ICE was performed in a patient undergoing ventricular tachycardia (VT) ablation. RESULTS: ICE showed extensive mobile thrombi on the implantable cardioverter defibrillator lead. Radiofrequency catheter ablation of VT from perimitral scar was safely performed via a retrograde aortic approach. After the procedure, chronic anticoagulation was initiated. CT-angiography of the chest 2 months later showed no pulmonary emboli. CONCLUSIONS: The significance of these thrombi, as related to chronic pulmonary embolization, warrants further study.


Subject(s)
Catheter Ablation , Heart Diseases , Tachycardia, Ventricular , Thrombosis , Humans , Tachycardia, Ventricular/diagnostic imaging , Tachycardia, Ventricular/surgery , Wind , Heart Diseases/surgery , Thrombosis/diagnostic imaging , Thrombosis/etiology , Catheter Ablation/methods , Hair , Treatment Outcome
10.
JACC Clin Electrophysiol ; 9(11): 2303-2314, 2023 11.
Article in English | MEDLINE | ID: mdl-37632506

ABSTRACT

BACKGROUND: Intramural hematoma during ablation for scar-related ventricular tachycardia (VT) is a rare but life-threatening complication. OBJECTIVES: The goal of this study was to describe the features and outcomes of intramural hematoma during ablation for scar-related VT. METHODS: From 2010 to 2022, >3,514 ablations for ventricular arrhythmias were performed at 2 institutions. Four cases of intramural hematoma complicating VT ablation for scar-related VT were identified. Intraprocedural details, imaging data, and surgical notes were reviewed to create a recognizable pattern of events highlighting this complication. RESULTS: In 3 of 4 cases, intramural hematoma occurred during catheter ablation with an open irrigated 3.5 mm tipped catheter using normal saline for irrigation. In one case, hematoma was noted after ablation using an investigational needle electrode catheter. The occurrence of a steam pop preceded detection of an expanding intramural hematoma in 3 cases. ST-segment elevation on electrocardiography was evident in 3 cases; intracardiac echocardiographic imaging detected the hematoma in all cases. Epicardial rupture and pericardial effusion requiring drainage occurred in 3 cases, whereas 1 hematoma was self-contained and did not require intervention. Surgical intervention was performed in 2 cases, with successful outcomes. One patient who was deemed not a surgical candidate died of progressive cardiogenic shock. CONCLUSIONS: Intramural hematoma during ablation for scar-related VT is a rare but potentially catastrophic complication that requires prompt recognition. Steam pops during ablation frequently precede the hematoma formation. Surgical intervention may be life-saving, although contained hematomas can occasionally be managed conservatively.


Subject(s)
Catheter Ablation , Tachycardia, Ventricular , Humans , Cicatrix/complications , Cicatrix/surgery , Cicatrix/pathology , Steam , Tachycardia, Ventricular/etiology , Tachycardia, Ventricular/surgery , Tachycardia, Ventricular/diagnosis , Catheter Ablation/adverse effects , Catheter Ablation/methods , Hematoma/diagnostic imaging , Hematoma/etiology , Hematoma/surgery
12.
JACC Clin Electrophysiol ; 9(7 Pt 2): 1147-1157, 2023 07.
Article in English | MEDLINE | ID: mdl-37495323

ABSTRACT

BACKGROUND: Experimental evidence suggests genetic variation in 4q25/PITX2 modulates pulmonary vein (PV) myocardial sleeve length. Although PV sleeves are the main target of atrial fibrillation (AF) ablation, little is known about the association between different PV sleeve characteristics with ablation outcomes. OBJECTIVES: This study sought to evaluate the association between clinical and genetic (4q25) risk factors with PV sleeve length in humans, and to evaluate the association between PV sleeve length and recurrence after AF ablation. METHODS: In a prospective, observational study of patients undergoing de novo AF ablation, PV sleeve length was measured using electroanatomic voltage mapping before ablation. The sentinel 4q25 AF susceptibility single nucleotide polymorphism, rs2200733, was genotyped. The primary analysis tested the association between clinical and genetic (4q25) risk factors with PV sleeve length using a multivariable linear regression model. Covariates included age, sex, body mass index, height, and persistent AF. The association between PV sleeve length and atrial arrhythmia recurrence (>30 seconds) was tested using a multivariable Cox proportional hazards model. RESULTS: Between 2014 and 2019, 197 participants were enrolled (median age 63 years [IQR: 55 to 70 years], 133 male [67.5%]). In multivariable modeling, men were found to have PV sleeves 2.94 mm longer than women (95% CI: 0.99-4.90 mm; P < 0.001). Sixty participants (30.5%) had one 4q25 risk allele and 6 (3.1%) had 2 alleles. There was no association between 4q25 genotype and PV sleeve length. Forty-six participants (23.4%) experienced arrhythmia recurrence within 3 to 12 months, but there was no association between recurrence and PV sleeve length. CONCLUSIONS: Common genetic variation at 4q25 was not associated with PV sleeve length and PV sleeve length was not associated with ablation outcomes. Men did have longer PV sleeves than women, but more research is needed to define the potential clinical significance of this observation.


Subject(s)
Atrial Fibrillation , Pulmonary Veins , Female , Humans , Male , Middle Aged , Atrial Fibrillation/genetics , Atrial Fibrillation/surgery , Genotype , Prospective Studies , Pulmonary Veins/surgery , Risk Factors , Aged , Homeobox Protein PITX2
13.
JACC Clin Electrophysiol ; 9(8 Pt 2): 1475-1486, 2023 08.
Article in English | MEDLINE | ID: mdl-37278684

ABSTRACT

BACKGROUND: We previously reported feasibility of irrigated needle ablation (INA) with a retractable 27-G end-hole needle catheter to treat nonendocardial ventricular arrhythmia substrate, an important cause of ablation failure. OBJECTIVES: The purpose of this study was to report outcomes and complications in our entire INA-treated population. METHODS: Patients with recurrent sustained monomorphic ventricular tachycardia (VT) or high-density premature ventricular contractions (PVCs) despite radiofrequency ablation were prospectively enrolled at 4 centers. Endpoints included a 70% decrease in VT frequency or PVC burden decrease to <5,000/24 h at 6 months. RESULTS: INA was performed in 111 patients (median: 2 failed prior ablations, 71% nonischemic heart disease, and left ventricular ejection fraction 36% ± 14%). INA acutely abolished targeted PVCs in 33 of 37 patients (89%), and PVCs were reduced to <5,000/day in 29 patients (78%). During 6-month follow-up, freedom from hospitalization was observed in 50 of 72 patients with VT (69%), and improvement or abolition of VT occurred in 47%. All patients received multiple INA applications, with more in the VT group than in the PVC group (median: 12 [IQR: 7-19] vs 7 [5-15]; P < 0.01). After INA, additional endocardial standard radiofrequency ablation was required in 23% of patients. Adverse events included 4 pericardial effusions (3.5%), 3 cases of (anticipated) atrioventricular block (2.6%), and 3 heart failure exacerbations (2.6%). During 6-month follow-up, 5 deaths occurred; none were procedure-related. CONCLUSIONS: INA achieves improved arrhythmia control in 78% of patients with PVCs and avoids hospitalization in 69% of patients with VT refractory to standard ablation at 6-month follow-up. Procedural risks are acceptable. (Intramural Needle Ablation for Ablation of Recurrent Ventricular Tachycardia, NCT01791543; Intramural Needle Ablation for the Treatment of Refractory Ventricular Arrhythmias, NCT03204981).


Subject(s)
Catheter Ablation , Tachycardia, Ventricular , Ventricular Premature Complexes , Humans , Catheter Ablation/adverse effects , Stroke Volume , Ventricular Function, Left
15.
J Cardiovasc Electrophysiol ; 34(3): 769-774, 2023 03.
Article in English | MEDLINE | ID: mdl-36682066

ABSTRACT

Most currently available pacing and defibrillation leads utilize a stylet-based design that facilitates implantation. This has advantages, but also increases the lead diameter and adds the potential for metal fatigued-based conductor failure. A systematic literature search was conducted, and the authors add their twenty-year experience with this lead design. The global experience with lumenless leads was reviewed both for "standard" positioning and with conduction system pacing. Methods for both placement and system modification are reviewed. Lumenless leads have the potential to improve the durability of endocardial pacing and facilitate conduction system pacing.


Subject(s)
Pacemaker, Artificial , Humans , Equipment Design , Cardiac Pacing, Artificial/methods , Prosthesis Implantation/methods , Cardiac Conduction System Disease
16.
Circ Arrhythm Electrophysiol ; 16(2): e010889, 2023 02.
Article in English | MEDLINE | ID: mdl-36602818

ABSTRACT

BACKGROUND: Initiation of ventricular tachycardia (VT) by programmed electrical stimulation (PES) has an important role to allow mapping and assess ablation end points. We hypothesized that substrate mapping may alter VT inducibility by mechanical bumping of critical sites. METHODS: Subjects with left ventricular scar-related VT that was inducible by PES who were undergoing ablation were included. PES was repeated after substrate mapping (Group I) or after time under sedation/anesthesia during which additional imaging and transeptal puncture were performed without substrate mapping (Group II). The response to the second PES was categorized as type I if the same VT was induced, type II if a different VT was induced, and type III if VT was not inducible. RESULTS: Twenty-eight patients (median age 66 years, 61% ischemic cardiomyopathy), 14 in Group I and 14 in Group II, were included. Age, time between initial and second PES, type of cardiomyopathy, ejection fraction, and anesthesia methods were not different between the 2 groups. Initial VT cycle length, however, was shorter in Group I (305 millisecond [range, 235-600] versus 350 millisecond [range, 235-600], P=0.016). Also, Group I required more extrastimuli to induce VT in PES 1 (2 [1-4] versus 2 [1-3], P=0.022). In Group I, following substrate mapping, the second PES induced the same VT in 3 patients (21%), a different VT in 9 (64%), and no VT in 2 (14%) patients. In contrast, in Group II the same VT was induced in 10 (71%) patients, a different VT in 3 (21%) and no VT in 1 (7%) patient (P=0.017). CONCLUSIONS: Mechanical effects of substrate mapping commonly alter inducibility of VT. This has important implications for catheter ablation procedure planning and acute assessment of outcome and can potentially account for some recurrent VTs that are not recognized at the time of the procedure.


Subject(s)
Anesthesia , Cardiomyopathies , Catheter Ablation , Myocardial Ischemia , Tachycardia, Ventricular , Humans , Aged , Cardiomyopathies/surgery , Catheter Ablation/methods , Treatment Outcome
19.
J Cardiovasc Electrophysiol ; 33(12): 2658-2662, 2022 12.
Article in English | MEDLINE | ID: mdl-36125446

ABSTRACT

INTRODUCTION: Leadless pacemakers represent an increasingly utilized alternative to traditional pacing methods in those with prior bacteremia or at high risk for infection. The acknowledged resistance to infection is illustrated by the exceedingly rare documentation of it. METHODS: We present a case of methicillin-sensitive Staphylococcus aureus endocarditis with associated leadless pacemaker infection necessitating percutaneous aspiration of the device-associated vegetation followed by extraction of the leadless pacemaker. RESULTS: Large vegetation associated with a leadless pacemaker was percutaneously aspirated with a vacuum-assisted aspiration device, followed by successful extraction of the leadless pacemaker. CONCLUSION: While leadless pacemakers are seldom involved in infective endocarditis, ultrasound evaluation in high-risk patients with an undetermined source is reasonable. Before extraction, it is practical to consider aspiration of large associated vegetations with a vacuum-assisted device.


Subject(s)
Endocarditis, Bacterial , Endocarditis , Pacemaker, Artificial , Humans , Treatment Outcome , Pacemaker, Artificial/adverse effects , Device Removal/methods , Endocarditis, Bacterial/diagnosis , Endocarditis, Bacterial/therapy , Postoperative Complications/etiology , Equipment Design