Opt Express ; 21(22): 25677-84, 2013 Nov 04.
Article in English | MEDLINE | ID: mdl-24216793

ABSTRACT

We have carried out grating-based x-ray differential phase-contrast measurements with a hybrid pixel detector in 16 energy channels simultaneously. A method for combining the energy resolved phase-contrast images based on energy weighting is presented. An improvement in contrast-to-noise ratio by 58.2% with respect to an emulated integrating detector could be observed in the final image. The same image quality could thus be achieved with this detector and with energy weighting at 60.0% reduced dose compared to an integrating detector. The benefit of the method depends on the object, spectrum, interferometer design and the detector efficiency.


Subject(s)
Photometry/instrumentation , Radiographic Image Interpretation, Computer-Assisted/instrumentation , Refractometry/instrumentation , Tomography, X-Ray Computed/instrumentation , Transducers , Equipment Design , Equipment Failure Analysis , Photons