Tex Heart Inst J ; 34(4): 406-11, 2007.
Article in English | MEDLINE | ID: mdl-18172519

ABSTRACT

In this study, long-term (90-day) hemocompatibility and end-organ effects of a centrifugal left ventricular assist device (the Heartware HVAD) were evaluated in 6 healthy sheep. The device was implanted into the left ventricular apex on beating hearts. The outflow graft of each device was anastomosed to the descending aorta. None of the sheep received anticoagulation or antiaggregation medication during the study. Hematologic and biochemical tests of liver and kidney function were performed pre-operatively (baseline) and throughout the study. Data associated with pump function were collected continuously until 90 +/- 1 days of support, at which time the sheep were humanely killed, and the end-organs were examined macroscopically and histopathologically. Hematologic and biochemical test results were within normal limits during the study period. There were no significant complications. Postmortem examination of the explanted organs revealed no evidence of ischemia or infarction, except in 2 sheep, in which small foci of infarction were detected in each of their left kidneys. There was no significant device failure. In all sheep, the pump's inflow and outflow conduits were free of thrombus. During the 90-day study, the HeartWare HVAD showed exceptional hemocompatibility and reliability, both of which are crucial to the clinical success of any implantable left ventricular assist device.


Subject(s)
Biocompatible Materials , Heart Failure/surgery , Heart-Assist Devices , Prosthesis Implantation/instrumentation , Animals , Disease Models, Animal , Erythrocyte Count , Follow-Up Studies , Heart Failure/blood , Heart Ventricles/pathology , Hematocrit , Prosthesis Design , Reproducibility of Results , Sheep
SELECTION OF CITATIONS
SEARCH DETAIL