J Pediatr Gastroenterol Nutr ; 42(3): 331-5, 2006 Mar.
Article in English | MEDLINE | ID: mdl-16540807

ABSTRACT

We evaluated the accuracy with which the Hologic QDR-4500A and QDR-2000 densitometers measure fat mass (FM) in 95 children. FM was derived from total body water measured by deuterium dilution (DD) in all children, by QDR-4500A in 50, and by QDR-2000 in 45 children. Compared with DD, both instruments underestimated FM (QDR-4500A: 3.35 +/- 2.5 kg, P < 0.0001; QDR-2000: 1.05 +/- 1.5 kg, P < 0.0001). Both absorptiometers showed magnitude biases relative to DD (QDR-4500: r = +0.70, P < 0.001; QDR-2000: r = -0.51, P < 0.001). We conclude that neither densitometer is equivalent to DD for estimation of children's FM. The QDR-4500A's current calibration seems to provide an even greater underestimate of FM than the QDR-2000.


Subject(s)
Absorptiometry, Photon/methods , Adipose Tissue , Body Composition/physiology , Deuterium , Absorptiometry, Photon/instrumentation , Absorptiometry, Photon/standards , Body Water/metabolism , Calibration , Child , Female , Humans , Male