Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters










Database
Language
Publication year range
1.
Biomed Instrum Technol ; 50(5): 336-48, 2016.
Article in English | MEDLINE | ID: mdl-27632039

ABSTRACT

A battery-operated active cooling/heating device was developed to maintain thermoregulation of trauma victims in austere environments while awaiting evacuation to a hospital for further treatment. The use of a thermal manikin was adopted for this study in order to simulate load testing and evaluate the performance of this novel portable active cooling/heating device for both continuous (external power source) and battery power. The performance of the portable body temperature conditioner (PBTC) was evaluated through cooling/heating fraction tests to analyze the heat transfer between a thermal manikin and circulating water blanket to show consistent performance while operating under battery power. For the cooling/heating fraction tests, the ambient temperature was set to 15°C ± 1°C (heating) and 30°C ± 1°C (cooling). The PBTC water temperature was set to 37°C for the heating mode tests and 15°C for the cooling mode tests. The results showed consistent performance of the PBTC in terms of cooling/heating capacity while operating under both continuous and battery power. The PBTC functioned as intended and shows promise as a portable warming/cooling device for operation in the field.


Subject(s)
Equipment Failure Analysis/instrumentation , Heating/instrumentation , Hyperthermia, Induced/instrumentation , Hypothermia/therapy , Manikins , Equipment Design , Humans , Hypothermia/diagnosis , Reproducibility of Results , Sensitivity and Specificity , Skin Temperature
SELECTION OF CITATIONS
SEARCH DETAIL
...