Pacing Clin Electrophysiol ; 36(2): e53-5, 2013 Feb.
Article in English | MEDLINE | ID: mdl-21535030

ABSTRACT

Occurrence of supraventricular tachycardia is a common cause of clinical impairment for patients implanted with CRT devices. We report the case of atrial activity oversensing by the left ventricular (LV) lead during typical flutter, which led to LV pacing inhibition. Temporary reprogramming of the LV detection from standard bipolar to extended bipolar and cavotricuspid isthmus ablation solved this problem.


Subject(s)
Atrial Flutter/diagnosis , Atrial Flutter/prevention & control , Electrocardiography/methods , Equipment Failure , Tachycardia, Supraventricular/diagnosis , Tachycardia, Supraventricular/prevention & control , Therapy, Computer-Assisted/methods , Aged , Electrodes, Implanted , Heart Ventricles , Humans , Male , Treatment Outcome