Phys Med Biol ; 60(17): 6669-83, 2015 Sep 07.
Article in English | MEDLINE | ID: mdl-26271097

ABSTRACT

In this work we use EBT3 film measurements at 10 MV to demonstrate the suitability of the Exradin W1 (plastic scintillator) for relative dosimetry within small photon fields. We then use the Exradin W1 to measure the small field correction factors required by two other detectors: the PTW unshielded Ediode 60017 and the PTW microDiamond 60019. We consider on-axis correction-factors for small fields collimated using MLCs for four different TrueBeam energies: 6 FFF, 6 MV, 10 FFF and 10 MV. We also investigate percentage depth dose and lateral profile perturbations. In addition to high-density effects from its silicon sensitive region, the Ediode exhibited a dose-rate dependence and its known over-response to low energy scatter was found to be greater for 6 FFF than 6 MV. For clinical centres without access to a W1 scintillator, we recommend the microDiamond over the Ediode and suggest that 'limits of usability', field sizes below which a detector introduces unacceptable errors, can form a practical alternative to small-field correction factors. For a dosimetric tolerance of 2% on-axis, the microDiamond might be utilised down to 10 mm and 15 mm field sizes for 6 MV and 10 MV, respectively.


Subject(s)
Photons , Scintillation Counting/methods , Algorithms , Scintillation Counting/instrumentation