Your browser doesn't support javascript.
loading
Show: 20 | 50 | 100
Results 1 - 1 de 1
Filter
Add more filters










Database
Language
Publication year range
1.
Med Dosim ; 46(4): 426-430, 2021.
Article in English | MEDLINE | ID: mdl-34167869

ABSTRACT

NRG-CC001 recently reported positive results on hippocampal-sparing IMRT (HS-IMRT) in conjunction with memantine for the reduction in cognitive decline compared to conventional whole brain radiation therapy. Herein, we report our experience in planning volumetric modulated arc therapy (VMAT) cases in Monaco® with the anticipation of increased utilization of the planning technique for delivery on Elekta linear accelerators. Twelve patients previously treated with whole brain radiation therapy who would have been eligible for NRG-CC001 were replanned with VMAT HS-IMRT for to a dose of 30Gy/10fx using constraints from the trial. All 12 patients were able to be planned with VMAT and achieve NRG-CC001 dose constraints. Median maximum and D100% to the right and left hippocampi were: 13.37Gy and 13.43Gy, respectively and 8.76Gy and 8.86Gy, respectively. Median coverage of the brain minus the hippocampi with 30Gy was 96.53%. All cases passed quality assurance testing with 3%/3 mm and 2%/2 mm criterion. Hippocampal-sparing IMRT whole brain radiation therapy can be feasibly planned with VMAT technique in Monaco® and delivered on Elekta linear accelerators.


Subject(s)
Radiotherapy, Intensity-Modulated , Hippocampus , Humans , Radiotherapy Dosage , Radiotherapy Planning, Computer-Assisted
SELECTION OF CITATIONS
SEARCH DETAIL
...