J Pediatr (Rio J) ; 76(6): 447-52, 2000.
Article in Portuguese | MEDLINE | ID: mdl-14647633

ABSTRACT

OBJECTIVE: To evaluate the accuracy of the Mini-Wright (Clement Clarke International Ltd.) peak-flow meters. METHODS: Twenty of those meters were checked by use of electronic calibration syringe (Jones Flow-Volume Calibrator(R)). Nine of them had an old scale, with values displayed equidistantly, and eleven had a new mechanical scale with non-equidistant values. Each device was connected in series to the calibration syringe to perform eight hand-driven volume injections, with flows ranging from 100 to 700 l/min. Absolute and relative differences between meters and syringe were calculated, the syringe values taken as standard. The accuracy of the twenty Mini-Wright devices was validated by the American Thoracic Society criteria (-/+ 10% or -/+ 20 l/min), and/or European Respiratory Society criteria (-/+ 5% or -/+ 5 l/ min). RESULTS: New scale instruments were more accurate than old scale meters (p < 0.001), by both ATS and ERS criteria. Every meter was rechecked after 600 measurements. Both the old, and the new scale instruments maintained the same level of performance after this evaluation. CONCLUSIONS: Results suggest that new scale meters were accurate and can be safely used in clinical practice. The authors strongly recommend that they are rechecked regularly to ensure that they are within the ATS and ERS variation limits.