Eur J Anaesthesiol ; 13(2): 143-6, 1996 Mar.
Article in English | MEDLINE | ID: mdl-8829947

ABSTRACT

The occurrence of a pneumothorax in a 4-week-old infant is described. The cause of the problem is believed to have been a kink in the silastic reservoir tube. The failure of the monitors to detect the event and methods of improving safety are discussed. Difficulties may occur even with familiar equipment; the value of full monitoring and constant vigilance are stressed.


Subject(s)
Pneumothorax/etiology , Respiration, Artificial/adverse effects , Humans , Infant, Newborn , Intraoperative Complications , Male , Monitoring, Intraoperative , Pneumothorax/diagnostic imaging , Radiography , Respiration, Artificial/instrumentation