1.
Appl Radiat Isot ; 70(9): 2063-5, 2012 Sep.
Article in English | MEDLINE | ID: mdl-22516716

ABSTRACT

A prototype of a calibration facility for noble gas monitoring using (41)Ar in the PTKMR-BATAN has been tested. The facility was designed in such a way that the standard source of gas can be reused. The radioactive (41)Ar source was obtained by thermal neutron reaction of (40)Ar(n, γ)(41)Ar using a thermal neutron flux of 4.8×10(13) neutrons per cm(2) per second in two minutes on the multipurpose G.A. Siwabessy Reactor (Batan, Serpong, Indonesia). Gamma spectrometry was used to measure the radioactivity and purity of (41)Ar. The spectrum of the (41)Ar observed yields an energy of 1294 keV because of the highest intensity (99.2%). The activity of (41)Ar was 2821 kBq and 4% of the expanded uncertainty. The time required for (41)Ar to reach homogeneity was 7 min, and the effectiveness of resuse was 53%.


Subject(s)
Argon/analysis , Argon/standards , Gases/standards , Radioisotopes/analysis , Radioisotopes/standards , Radiometry/instrumentation , Radiometry/standards , Calibration , Equipment Design , Equipment Failure Analysis , Gases/analysis , Half-Life , Internationality , Radiation Dosage , Reference Standards , Reference Values